Instruction     Manual 

                 

           

             

 

 

 

 

 

 

IVC2300

 

VACUUM GUAGE CONTROLLER 

 

 

 

 

 

 

 

 

 

 

 

 

 

Address : #302, 608-2, Bora-dong, Giheung-gu,  

Yongin-si, Kyeonggido, Korea 

Tel : +82-31-284-0791 

Fax : +82-31-284-0792 

Home Page : www.isvt.co.kr  

E-mail : [email protected] 

Содержание IVC2300

Страница 1: ...struction Manual IVC2300 VACUUM GUAGE CONTROLLER Address 302 608 2 Bora dong Giheung gu Yongin si Kyeonggido Korea Tel 82 31 284 0791 Fax 82 31 284 0792 Home Page www isvt co kr E mail jbkim isvt co k...

Страница 2: ...WARRANTY Vacuum Gauge Controller IVC2300 1 VC121 SENSOR Interface Demo Program Internet Home Page http www isvt co kr...

Страница 3: ...Grounding Manual CAUTION IVC2300 Vacuum Gauge Controller 195 265VAC 50 60Hz 110VAC SAFETY WARNING Gas Gas Controller Sensor Wire Sensor Wire Gas NOTE Manual Tel 031 284 0791 web site http www isvt co...

Страница 4: ...Customer Support SENSER IVC2300 Vacuum Gauge Controller 608 2 302 2...

Страница 5: ...0 Ionization Gauge Dimension Pin Out 11 IVC2300 Controller Dimension Mounting 12 Electrical Information Connector Pin Out Convection Gauge Connector 13 Capacitance Manometer CM Connector Ionization Ga...

Страница 6: ...Connecting Cable Gauge Mounting Position Convection Gauge Ionization Gauge Cable Connection Convection Gauge 9 Pin D Sub Connector Ionization Gauge Cable IVC2300 Vacuum Gauge Controller Flash Memory T...

Страница 7: ...Independently Set SPDT Contact 0 5 Amp 125 VAC Resistive Load Pressure Voltage Output D A Output 0 10 VDC Measuring Range Convection Gauge 1000 8 0 X 10 4 Torr Ionization Gauge 1 0X10 3 1 0X10 9Torr...

Страница 8: ...Series 4336 and any Bayard Alpert Gauges Mounting Position Vertical Envelope Nonex 7720 Glass Collector 0 005 Tungsten Filament Thoria Coated Iridium or Tungsten Grid Tungsten Helix Configuration Bak...

Страница 9: ...npack Unpacking Check List IVC2300 Instruction Manual this book IVC2300 Control Unit VC121 Series Convection Gauge ETI Series 4336 Ionization Gauge Convection Gauge Cable Ionization Gauge Cable Power...

Страница 10: ...Connecting Cables Gauge Mounting PC RS232 IVC2300 VC120 VC120 CM IG Figure 1 Connecting Cable Gauge Mounting 8...

Страница 11: ...Installation Environment Operating Environmental Requirements 15 50 59 122 Chassis Grounding 9...

Страница 12: ...Installation Convection Gauge Dimension Pin out Figure 2 Convection Gauge Dimension NW16KF Fitting Option CF Flange 10...

Страница 13: ...Grid White Filament Yellow Filament Red Grid Black Ionization Gauge Dimension Pin out Figure 3 Ionization Gauge Dimension Pin out 11...

Страница 14: ...IVC2300 Controller Dimension Mounting Figure 4 Front Panel Dimension Figure 5 Cutting Size Dimension 12 W W W I S V T C O K R LABEL SP4 IG SP4 SP4 SP4 SP4 SP4 ON OFF ON OFF DEGAS MENU ESC ENT...

Страница 15: ...ved Pin Convection Gauge CG A CG B Connector 9 Pin Female D sub Type Pin Assignment Convection Gauge Connector Pin out Pin No Assignment Cable Color 1 Drive Bridge Brown 2 Signal Ground Red 3 Bridge H...

Страница 16: ...ly White 6 No connection 7 Ground Signal Brown 8 Ground Signal Black 9 No Connection Table 4 Capacitance Manometer Connector Pin out Ionization Gauge IG Connector Ionization Gauge Connector 4 Pin Male...

Страница 17: ...in Assignment Serial Interface RS 232 Connector Pin out Pin No Assignment Cable Color 1 No Connection 2 Transmit Data Orange 3 Received Data Red 4 No Connection 5 Digital Ground Black 6 No Connection...

Страница 18: ...P 4 Common 8 S P 4 Normal Open 9 S P 5 Common 10 S P 5 Normal Open 11 S P 6 Common 12 S P 6 Normal Open 13 CM Analog Output 14 CG B Analog Output 15 CG A Analog Output 16 IG Analog Output 17 Analog Gr...

Страница 19: ...apter Operation Front Panel Components IVC2300 Table 8 Display Control Key Trip Point Relay LCD User Set Point 17 LABEL SP4 IG SP4 SP4 SP4 SP4 SP4 ON OFF ON OFF DEGAS MENU ESC ENT W W W I S V T C O K...

Страница 20: ...B Capacitance Manometer CM Display Trip Point Relay Label Window User Set Point IG ON OFF Key Key Ionization Gauge 10 3 High Level Gauge OFF Gauge Degas key Ionization Gauge Degas Key Degas Time 30 D...

Страница 21: ...n Curr ent Default 1 0mA Setting 2 Degas Time IG Degas Set 30 sec Degas Time 30 Default Setting 3 Auto IG Ref Auto IG Ref Set 1 CG Channel A 2 CG Channel B 3 Auto IG Off Convection Gauge IG On Off Def...

Страница 22: ...Zero CG A Zero CAL 10 4 Convection Gauge Zero 4 CG B Zero CG B Zero CAL 10 4 Convection Gauge Zero 5 Zero CGA Init CG A Initaialize Zero Calibration 6 Zero CGB Init CG B Initaialize Zero Calibration 7...

Страница 23: ...ain Menu Sub Menu Descriptions 3 CM Set 1 CM Range CM Range Set 1000 Torr CM Gauge Range 2 CM Zero CM Zero Cal CM Zero CM Gauge Resolution CM Gauge Zero 3 To Upper Menu Menu 4 Exit Pressure Display Ta...

Страница 24: ...e 1 IG 2 CG Channel A 3 CG Channel B 4 CM Trip Point Relay 6 Gauge Setting 3 SP3 Ref Gauge SP3 Ref Gauge 1 IG 2 CG Channel A 3 CG Channel B 4 CM Trip Point Relay 6 Gauge Setting 4 SP4 Ref Gauge SP4 Re...

Страница 25: ...ip Point Relay 6 set point 3 SP3 Set Setpoint 3 SET 0 0 X 10E 0Torr Trip Point Relay 6 set point 4 SP4 Set Setpoint 4 SET 0 0 X 10E 0Torr Trip Point Relay 6 set point 5 SP5 Set Setpoint 5 SET 0 0 X 10...

Страница 26: ...CD 7 SYS Set Initialize 1 Sys CTRL Mode Sys CTRL Mode 1 Local 2 TTL 3 RS232 4 D Net Not Yet Controller Local Interface setting 2 Comm BAUD Rate Comm Baud Rate 9600 14400 28800 57600 Computer 4 To Uppe...

Страница 27: ...Analog Output Formula IG Vout V log P 12 P 10 V 12 CG Vout V log P 7 P 10 V 7 25...

Страница 28: ...Character R means READ and S means SET W means WRITE C means Calibration READ Command Return Value 9byte data RDI Read ION Pressure Data RDA Read CGA Pressure Data RDB Read CGB Pressure Data RS1 Read...

Страница 29: ...ION 2 CGA 3 CGB Therefor SP Ref1 2 is ION SP Ref3 4 is CGA SP Ref5 6 is CGB RSS Read Set point Status LED Status Return Value 6byte data EX Host RSS CR IV2300 110110 Parameter means that 6BYTE Date is...

Страница 30: ...Value 1byte data EX Host RIC CR IV2300 1 Parameter means that 1 0 1mm 2 1mm 3 10mm RIS Read IG Status Return Value 1byte data EX Host RIS CR IV2300 1 Parameter means that 1 IG ON 0 IG OFF SAR Set Auto...

Страница 31: ...ON SDE Set DEGAS ON OFF Return Value OK or INVALID Parameter 0 or 1 Parameter means that 0 OFF 1 ON EX Host SDE 1 CR IV23000 OK It is DEGAS ON SDT Set DEGAS Time Return Value OK or INVALID Parameter...

Страница 32: ...m SLB Set LCD Brightness Return Value OK or INVALID Parameter 1 16 Parameter means that LCD Brightness value EX Host SLB 10 CR IV2300 OK It is LCD Brightness setting by 10 CZA CAL ZERO CGA Return Valu...

Страница 33: ...CZB CAL ZERO CGB Return Value OK Parameter None EX Host CZB CR IV2300 OK It is CGB ZERO Calibration CAB CAL ATM CGB Return Value OK Parameter None EX Host CAB CR IV2300 OK It is CGB ATM Calibration CI...

Страница 34: ...urn Value OK Parameter 1 3 Parameter means that Ref Setting to 1 IG 2 CGA 3 CGB If CM Card is used instead IG Card 1 is CM EX Host SR1 2 CR IV2300 OK It is setting SP1 ref by CGA SR2 Set Setpoint Ref2...

Страница 35: ...t Setpoint Ref6 Return Value OK Parameter 1 3 EX Host SR6 2 CR IV2300 OK It is setting SP6 ref by CGA SRA Set Ref All SR1 SR6 Return Value OK Parameter 1 3 6Byte Data Parameter means that Ref Setting...

Страница 36: ...Value OK or INVALID Refer to IV2300_RS232_SP_TABLE EX Host WS1 150 CR IV2300 OK If SP1 Ref is IG SP1 value is 4 0 10E 4 If SP1 Ref is CGA or CGB SP1 value is 6 8 10E 3 WAS Write Auto IG SP Data Parame...

Страница 37: ...RCR Read CM Range return value 5byte data EX Host RCR CR IV2300 1000 SCR Set CM Range Return Value OK Parameter 1 3 Parameter means that Ref Setting to 1 15 means that 1 2 5 10 20 50 100 200 500 1000...