Water Quality Measuring Instrument
13
CHAPTER 3.
Trouble and Solution
Error
Reason
Solution
Low battery
Low battery
Change for new battery.
Power-
intensive
Battery’s bad quality
Change battery
Too many functions were used
Adjust to the suitable back light,
auto power off time, and the
voice according to customer
need.
Out of range
Turbidity is too high
Dilute the water sample
Please close the shielded cap
Cap was open.
Close the cap
Calibration failed
-Standard solution is not the
same as the set concentration.
-The deviation between
standard solution and
theoretical value is high.
-Sample cell is dirty or with
bubbles.
-Sample is not fully mixed.
Check the reasons and remove it.
Wrong password
Wrong password
Input correct password
The result is not accurate
Don’t read manual carefully
Read manual carefully.
Inappropriate sample
The sample must be uniform,
representative and sample should
be fully mixed or shake.
Fingerprints or water marks on
Wall of sample cell
Clean the cell with wiper
Testing time is too long or short
The sample need shake
before
[p1]
testing, and make
sure there are no bubbles, then
test the sample quickly.
Not boot
- Low battery.
- The backlight is too dark.
- Device was turned off without
operation over the set time
- Device error
- Change new battery
- Setting the backlight
- Press any key to boot
- Return to seller for repair
3. Trouble and solution