background image

Specification

Model

Temperature Range*

Set Point Resolution

Accuracy

Stability

Time to Set Point

Additional Time

for best stability

Well dimensions

Power

Fuse Rating

Size

Weight

Certification

Quick Cal

Low Temperature

-12°C to +140°C**

0.1°C over range

±0.4°C to ±0.1°C using

Comparison Techniques

±0.05°C

9 minutes from 0°C to 100°C

Typically

5 minutes

2 x 13mm diameter

by 120mm deep

12V DC or

100V - 240V

50/60Hz

5A Anti-surge (T) 5 x 20mm

65 x 152 x 175mm

1.5 kg (3.3lbs)

Traceable Certification included 

NAMAS Certification available

Quick Cal
High Temperature

+30°C to +350°C

0.1°C over range

±0.4°C to ±0.1°C using
Comparison Techniques

±0.05°C

9 minutes from Ambient to 350°C

Typically
5 minutes

1 x 25 mm diameter
by 120mm deep

100-130V or
200-250V AC
50/60Hz

1.6A Quick Acting (F) 5 x 20mm

65 x 152 x 175mm

1.5 kg (3.3lbs)

Traceable Certification included
NAMAS Certification available

* In an ambient of 20°C.

The Quick Cal Low's minimum operating temperature is ambient dependant.

** Upper Temperature range is decreased when powered by a 12V D.C. Battery.

Accessories

Item

High

Low

Carry Case

931-22-52

931-22-52

Power Supply

n/a

935-42-20

Battery & Charger

n/a

935-42-14

Inserts

Blank Inserts

550-02-04A

560-02-08

Custom Drilled

Consult Factory

Consult Factory

The company is always willing to give technical advice 
and assistance where appropriate. Equally, because 
of the programme of continual development and 
improvement we reserve the right to amend or alter 
characteristics and design without prior notice. This 
publication is for information only.

The apparatus Quick-Cal High and Quick-Cal Low's 
optional power supply are provided with an approved 
power cord. If the plug is not suitable for your location 
then the plug should be removed and replaced with an 
appropriate plug.

Take care to ensure the old plug is disposed safely. 
The cable is colour coded as follows:

COLOUR

Green / Yellow

Brown

Blue

FUNCTION

Earth (Ground)

Live (Line)

Neutral

Please ensure that your unit is correctly connected to 
the electricity supply.

THE APPARATUS MUST BE CORRECTLY 

EARTHED (GROUNDED).

The units on/off switch is located on the rear panel. 
Take care NOT to switch the unit off when it is hot - 
allow to cool first.