Specification
Model
Temperature Range*
Set Point Resolution
Accuracy
Stability
Time to Set Point
Additional Time
for best stability
Well dimensions
Power
Fuse Rating
Size
Weight
Certification
Quick Cal
Low Temperature
-12°C to +140°C**
0.1°C over range
±0.4°C to ±0.1°C using
Comparison Techniques
±0.05°C
9 minutes from 0°C to 100°C
Typically
5 minutes
2 x 13mm diameter
by 120mm deep
12V DC or
100V - 240V
50/60Hz
5A Anti-surge (T) 5 x 20mm
65 x 152 x 175mm
1.5 kg (3.3lbs)
Traceable Certification included
NAMAS Certification available
Quick Cal
High Temperature
+30°C to +350°C
0.1°C over range
±0.4°C to ±0.1°C using
Comparison Techniques
±0.05°C
9 minutes from Ambient to 350°C
Typically
5 minutes
1 x 25 mm diameter
by 120mm deep
100-130V or
200-250V AC
50/60Hz
1.6A Quick Acting (F) 5 x 20mm
65 x 152 x 175mm
1.5 kg (3.3lbs)
Traceable Certification included
NAMAS Certification available
* In an ambient of 20°C.
The Quick Cal Low's minimum operating temperature is ambient dependant.
** Upper Temperature range is decreased when powered by a 12V D.C. Battery.
Accessories
Item
High
Low
Carry Case
931-22-52
931-22-52
Power Supply
n/a
935-42-20
Battery & Charger
n/a
935-42-14
Inserts
Blank Inserts
550-02-04A
560-02-08
Custom Drilled
Consult Factory
Consult Factory
The company is always willing to give technical advice
and assistance where appropriate. Equally, because
of the programme of continual development and
improvement we reserve the right to amend or alter
characteristics and design without prior notice. This
publication is for information only.
The apparatus Quick-Cal High and Quick-Cal Low's
optional power supply are provided with an approved
power cord. If the plug is not suitable for your location
then the plug should be removed and replaced with an
appropriate plug.
Take care to ensure the old plug is disposed safely.
The cable is colour coded as follows:
COLOUR
Green / Yellow
Brown
Blue
FUNCTION
Earth (Ground)
Live (Line)
Neutral
Please ensure that your unit is correctly connected to
the electricity supply.
THE APPARATUS MUST BE CORRECTLY
EARTHED (GROUNDED).
The units on/off switch is located on the rear panel.
Take care NOT to switch the unit off when it is hot -
allow to cool first.