The Quick-Cal range are designed to be fast, simple 
to use and above all truly portable. Ideal for the simple 
testing of temperature sensors.

OPERATION

Altering the Block Temperature, (Setpoint)

1.

Switch the unit on. The power switch is located 
on the rear panel.

2.

The controller will briefly show its software 
version before displaying an indication of the 
block temperature.

3.

Momentarily press either the UP or DOWN 
key once to display the setpoint (desired 
temperature).

4.

To alter the value press and hold the UP key to 
raise the value or the DOWN key to lower the 
value.

5.

The display will return to show the nominal 
block temperature when no key is pressed for 
0.5 second.

The other controller functions are hidden from the 
operator. The values are pre-set and should not be 
changed.

Getting Better Accuracy

For accurate calibration of sensors it is good practice 
to include a calibrated reference probe in the block, 
then the units under test are compared to this 
reference probe.

A B

100.0

A

'DOWN' Key

B

'UP' Key

Altering the Ramp Rate

The ramp rate or set point rate limit provides a means 
of limiting the rate at which the temperature of the heat 
source will increase or decrease. 

Normally it is desirable to allow the unit to heat and 
cool at the maximum rate possible, which is achieved 
by setting the ramp rate to OFF. This is the Factory set 
default value. 

For some applications, e.g. testing   temperature cut 
out switches it can be useful to be able to program the 
temperature ramp rate.

To alter the ramp rate:

1)

Press the SCROLL key until Sprr is displayed.

2)

The parameter can be changed from OFF to 
the desired value with the UP and DOWN 
keys.

3)

Press the SCROLL key to return to the normal 
display.

The units are °C per minute.

UNIT UNDER TEST

QUICK-CAL

REFERENCE

CALIBRATION

CERTIFICATE

TH

ISSUE 3, 12

 MAY 2000