Medusa 510/511 Series Iss.11 – 07/12
INITIAL TESTING
This unit was fully tested before despatch to you but please check its operation as outlined below.
After connecting the Medusa to the electricity supply, the temperature controller display will show the temperature
of the block and the last set-point value. The controller and indicator both go through a self-test sequence first.
Change the set-point to 100°C and observe that the block temperature rises and settles to this value. Place a
thermometer in the insert in the block and connect it to the suitably configured indicator. Confirm that the indicator
agrees within ±2°C of the controller.
Your unit should have performed as described above and can now be used for calibration.
If any problems or faults arise during these tests please contact us or our agents for help and advice.
IMPORTANT NOTICE
The controller's function settings are preset and will not require adjustment.