ISOMET RFA1170-4+iMS4-L Скачать руководство пользователя страница 4

                              

                               

 

  

ISOMET 

3. 

INSTALLATION AND ADJUSTMENT 

 

3.1 

For continuous or duty cycled operation above >2%, always connect cooling water. 

Flow rate greater than 2 litres / minute. Temperature < 20 deg.C  

(Factory fitted water connections suit 8mm OD plastic tubing). 

 

3.2 

C24V DC to the screw terminals of the filtered DC connector.  DO NOT APPLY 

POWER.  The DC supply should be capable of 40A 

 

 

3.2 

Connect the (4) RF output TNC jacks to the (4) RF inputs of the acousto-optic deflector (or a 

50

 RF high power load). The connection order depends on the AO Bragg orientation (see 

Figure 4) and will be either 1

1, 2

2, 3

3, 4

4 or 1

4, 2

3, 3

2, 4

1.  The deflector 

will not be damaged if the order is incorrect but the efficiency will be non-uniform. The 

amplifier outputs must be terminated. 

 

The cable lengths from the amplifier to the RF connections of the deflector must be equal. 

 

 

3.4 

Connect the Interlock of the acousto-optic deflector (7-way circular connector) to the 7-way 

connector J9 of the RFA1170/4. A suitable cable is supplied.  

 

The interlock signal triggers a fault conditionif the temperature of the modulator or internal 

driver temperature exceeds 30ºC. This will disable the RF outputs.  

 

There are two LED indicators 

1: 

The bulkhead LED (above RF output J2) is bi-colour. 

 

RED indicates 24Vdc is ON but the thermal interlocks and/or -ENABLE signal are 

inactive. 

 

Yellow indicates interlocks are OK and the -ENABLE signal is active. 

 

RF will be output as soon as the iMS4-L is programmed and initiated. 

 

2: 

LED stack on the iMS4-Linterface PCB  

A GREEN (or RED) LED will beat at 1Hz indicating that the interface control circuitry has 

initialized correctly  

 

 

DO NOT connect any of the interlock signal wires to chassis ground or 0V

 

 

 

Содержание RFA1170-4+iMS4-L

Страница 1: ...OT OPERATE the RFA1170 4 types or LS700 type AO DEVICES WITHOUT LIQUID COOLING Always refer to the AO test data sheet ISOMET CORP 5263 Port Royal Rd Springfield VA 22151 USA Tel 703 321 8301 Fax 703 3...

Страница 2: ...r to the AO device data sheet The RFA1170 4 iMS4 L combo includes a number of features to aid monitoring and control Temperature monitoring and thermal Interlock for the Amplifier and associated AO de...

Страница 3: ...puts Each channel features a class A B power stage to generate the final output level of up to 90W per output Water cooling is required The heatsink temperature must not exceed 70 C SERIOUS DAMAGE TO...

Страница 4: ...erminated The cable lengths from the amplifier to the RF connections of the deflector must be equal 3 4 Connect the Interlock of the acousto optic deflector 7 way circular connector to the 7 way conne...

Страница 5: ...n excess of this optimum level will cause a decrease in first order intensity a false indication of insufficient RF power and makes accurate Bragg alignment difficult It is therefore recommended that...

Страница 6: ...lue ISAT Typical values Class Method description Setting UpdateDDSPowerLevel 80 UpdateRFAmplitude Wiper 60 3 15 To equalise deflection efficiency across the extremes of the scan alternate between the...

Страница 7: ...tive RF output The Enable may remain LOW continuously The RF amplifiers are design to be biased at 100 duty factor T_DLY 1 8msec Pulsed Response All versions of the RA1170 4 are designed for fast puls...

Страница 8: ...ed with a soft ball of brushed short fibres removed cotton slightly moistened with clean alcohol Before the alcohol has had time to dry on the surface wipe again with dry cotton in a smooth continuous...

Страница 9: ...the optical surface and in the plane of deflection is Bragg fc 2 v The separation angle between the zeroth order and mid scan point of the first order is sep fc v The first order scan angle is scan f...

Страница 10: ...CLK TRG USB II Outline Connector Summary J1 TNC RF Output Ch1 J2 TNC RF Output Ch2 J3 TNC RF Output Ch3 J4 TNC RF Output Ch4 INT 7 way Circular Panel Plug Binder 712 AOD Temp and isolated Interlock I...

Страница 11: ...5 2 1 4 1 o H iD B 37 1 1 6 3 3 4 4 1 5 3 0 P AS ta g e T N C J1 O u tp u t P re A m p G a inS e t F P G A G a te T N C J2 O u tp u t T N C J3 O u tp u t T N C J4 O u tp u t D D S M C X M C X iM S 4 L...

Страница 12: ...Vdc 40A LED indicators 2 off 1 The bulkhead dual colour LED fitted above RF output J2 RED indicates 24Vdc is ON but one or more of the following conditions exist Interlock failure over temp ENABLE sig...

Страница 13: ...ay differ Connection options for Beam Steered AO Deflectors Amplifier Amplifier Amplifier RF1 RF4 J1 J3 J2 J4 J1 J3 J2 J4 Increasing Phase Delay Increasing Phase Delay J1 J3 J2 J4 Increasing Phase Del...

Отзывы: