MENU 6 - INPUTS
When one of this function is enabled, the related totalizer - may be reset through the on/off input.
Totalizers counting lock command enable. When this function is active, applying a voltage on the on/off input terminals the
system stops the totalizers no matter which is the flow rate.
When this function is active (ON), applying a voltage on the on input terminals, the measurement is stopped, the meter
will display zero flow.
When this function is active, applying a voltage on the on/off input terminals the meter performs a autozero calibration
cycle. ATTENTION: If the voltage pulse is less than 1 sec., the meter performs a calibration cycle to compensate possible
thermal drifts. If the voltage pulse is more 1 sec, the meter performs a zero calibration measure. To perform the calibration
it is absolutely necessary for the sensor to be full of liquid and that the liquid is perfectly still. Even very small movement
of the liquid may affect the result of the calibration, and, consequently, the accuracy of the system.
Range change external command enable. When this function is enabled, applying a voltage on the on/off input terminals
the meter switches to the second measuring range (Fs2).
N.B.: the autorange doesn’t allow to use the manual range change see (POS. 4.4 see page 34).
(POS. 6.3-4) Volume Totalizer Total /Partial Negative reset Enable
[T/P-/RESET=ON/OFF]
AL3
[VTTNE] [VTPNE]
(POS. 6.5) Totalizers Count Lock Input Enable
[COUNT LOCK= ON/OFF]
AL3
[TCLIE]
When one of this function is enabled, the related tot may be reset through the on/off input.
(POS. 6.1-2) Volume Totalizer Total /Partial Positive reset Enable
[T/P+/RESET=ON/OFF]
AL3
[VTTPE] [VTPPE]
(POS. 6.6) MeaSure Lock Input Enable
[Meas.lock=ON/OFF]
AL3
[MSLIE]
(POS. 6.7) CALibration Input Enable
[Calibration=ON/OFF]
AL3
[CALIE]
(POS. 6.8) Scale Range Change Input Enable
[RANGE CHANGE=ON/OFF]
AL3
[SRCIE]