![Iso-Tech SLM 52N Скачать руководство пользователя страница 5](http://html1.mh-extra.com/html/iso-tech/slm-52n/slm-52n_instruction-manual_2093341005.webp)
EN-3
Measurement functions:
•
Main processing functions
Sound level: Current time-weighted sound pressure level A or current time-
weighted sound pressure level C
Maximum time-weighted sound pressure level A or Maximum time-weighted
sound pressure level C
•
Total range: 30 to 130dB A & C weighted
•
Max. measurement level: 130dB
•
Self-generated noise level:
Typical values at 23
℃
using the nominal microphone equivalent capacitance of 27pF
(30-90dB range)
Weighting
Electrical
Total
“A”
22.7dB
26.1dB
“C”
21.8dB
29.5dB
Linearity operating range: A-weighted, 1000Hz, 60dB dynamic range.
Total linear operating range:
In accordance with IEC 61672-1, A-weighted, 1000Hz: 30dB to 130dB.
Level range selection:
5 ranges in 10dB steps 30 to 90dB , 40 to 100dB
50 to 110dB , 60 to 120dB
70 to 130dB
LINEAR OPERATING RANGES (L.O.R.)
RANGE: 30 – 90 dB. Test starting point 64 dB for all weightings and frequencies except
31.5Hz A-weighted, for which the starting point is 44 dB.
FREQUENCY
Hz
WEIGHTING
L.O.R.
dB
WEIGHTING
L.O.R.
dB
31.5
A
36.1 – 50.6
C
39.5 – 87.0
1000
A
36.1– 90.0
C
39.5 – 90.0
4000
A
36.1 – 90.0
C
39.5 – 89.2
8000
A
36.1 – 88.9
C
39.5 – 87.0
RANGE: 40 – 100 dB. Test starting point 74 dB for all weightings and frequencies except
31.5Hz A-weighted, for which the starting point is 54 dB.
FREQUENCY
Hz
WEIGHTING
L.O.R.,
dB
WEIGHTING
L.O.R
dB
31.5
A
40.0 – 60.6
C
40.0 – 97.0
1000
A
40.0 – 100.0
C
40.0 – 100.0
4000
A
40.0 – 100.0
C
40.0 – 99.2
8000
A
40.0 – 98.9
C
40.0 – 97.0