Iso-Tech SLM 52N Скачать руководство пользователя страница 23

EN-21

The human perception of a sound depends not only on the sound pressure level, but
also on the frequency. At high or low frequencies, a sound is felt to be less loud than a
sound of equal level in the midrange. The frequency weighting A compensates for this
effect and produces measurement results which are close to the perceived sound level.
For this reason, this type of frequency weighting is widely used for purposes such as
sound level evaluation.

The frequency weighting C curve produces almost flat response, but with a roll off below
31.5Hz and above 8000Hz. This is suitable for sound pressure level measurements in
situations with unwanted low-frequency or high-frequency components.

18. APPENDIX B RMS DETECTION CIRCUIT AND TIME WEIGHTING

The sound level meter uses rms detection. The effective value E (rms) is defined by the
following equation.

E rms

T

e dt

T

(

)

=

z

1

2

0

The voltage e which changes over time is raised to the power of 2, and integration for
the  time  interval  T  is  performed.  The  result  is  divided  by  T  and  the  square  root  is
extracted. The circuit configuration for performing the above mathematical operation
looks as follows.

ei 

      

                     

Power of 2  

ei

2

         

1

0

T

T

z

     

ei

2

   

 

  

ei

2

       

 

E(rms)

Input Voltage                                                                                              Output Voltage

During sound level measurements, the level often fluctuates drastically, which would
make it difficult to evaluate readings if some kind of averaging were not applied. Sound
level  meters  therefore  provide  the  capability  for  index  weighting  (index  averaging)
using  the  rms  circuit.  The  parameters  of  this  weighting  process  are  called  the  time
weightings, determined by the time constant (see next page).

Sound level meters usually have a F(Fast) and S(Slow) setting for the time weighting.
The time range that is considered for averaging is narrow in the F(Fast) setting and
wide in the S(Slow) setting. In the F(Fast) setting, the instantaneous level has a larger
bearing on the displayed value than in the S(Slow) setting. From the point of view of the
measurement objective, the F(Fast) setting is more suitable to situations  with swiftly
changing  sound  level,  whereas  the  S(Slow)  setting  yields  a  more  broadly  averaged
picture. The F(Fast) setting is more commonly used, and sound pressure level values
given without other indication are usually made with F(Fast) characteristics.

Time weightings and time constant

Time constant

Time

Weightings

Rise time

Decay time

F(Fast)

125ms

125ms

S(Slow)

1s

1s

Содержание SLM 52N

Страница 1: ...Instruction Manual SLM 52N Sound Level Meter FR EN IT DE ES Un Ov...

Страница 2: ...6 12 ADJUSTMENT DATA FOR CALIBRATOR B K TYPE 4226 PRESSURE MODE 16 13 TYPICAL INSTRUMENT FREQUENCY RESPONSE AT 0 INCIDENCE 16 14 TYPICAL FREQUENCY RESPONSE DUE TO CASE REFLECTIONS AT 0 INCIDENCE 17 15...

Страница 3: ...ht during storage and use Protect the instrument from air with high salt or sulphur content gases and stored chemicals as this may damage the delicate microphone and sensitive electronics Always turn...

Страница 4: ...t suitable for use with a frequency analyzer level recorder FFT analyzer graphic recorder etc 3 MEASUREMENT PARAMETERS The following parameters are used on the instrument LA A frequency weighting soun...

Страница 5: ...e with IEC 61672 1 A weighted 1000Hz 30dB to 130dB Level range selection 5 ranges in 10dB steps 30 to 90dB 40 to 100dB 50 to 110dB 60 to 120dB 70 to 130dB LINEAR OPERATING RANGES L O R RANGE 30 90 dB...

Страница 6: ...0 1000 A 60 0 120 0 C 60 0 120 0 4000 A 60 0 120 0 C 60 0 119 2 8000 A 60 0 118 9 C 60 0 117 0 RANGE 70 130 dB Test starting point 104 dB for all weightings and frequencies except 31 5Hz A weighted f...

Страница 7: ...vel for the diffuse field 94 0dB Frequency for acoustic testing 8000Hz Warm up time 2min Sampling interval Bar graph indication 125 ms approx Numeric indication 1 sec approx Microphone equivalent elec...

Страница 8: ...50 100 200 500 1000 2000 5000 10000 20000 Hz Freq units 31 5 Hz 63 Hz 125 Hz 250 Hz 500 Hz 1000 Hz 2000 Hz 4000 Hz 8000 Hz dB 0 1 0 4 0 0 0 5 0 2 0 0 0 2 0 0 0 5 Typical diffuse field response for ran...

Страница 9: ...d frequency weighting Output voltage 1Vrms at full scale of the range Output impedance 5k Load impedance 1M DC output Output voltage 10mV dB Output impedance 5k Load impedance 1M Power requirements On...

Страница 10: ...m for 80 modulation at 1KHz IEC 61672 1 Instrumentation standard classification group X and performance class 2 sound level meter No permanent degradation of performance loss of function change of op...

Страница 11: ...shows the sound level as a numeric value and a bar graph The display also shows the operation mode of the instrument the selected measurement parameters warning indications 3 Button Turns the meter o...

Страница 12: ...ns select the memory address to be displayed 10 DC AC output AC output signal with frequency weighting DC output signal corresponding to sound level 11 External DC 9V power supply jack Type 1 3 coaxia...

Страница 13: ...A C A Frequency weighting or C Frequency weighting indicator 6 Ov Over range indicator for processed value 7 Ov Over range indicator 8 Battery capacity indicator 9 FAST Fast time weighting indicator...

Страница 14: ...batteries press the button to turn off the instrument Remove the cover of the battery compartment Insert the new 9V battery Replace the battery cover Press the button to turn on the instrument and che...

Страница 15: ...ilar wind noise and strong air movements at the microphone can cause measurement errors Such effect can be reduced by using the windscreen Windscreen 4 Tripod Mounting For long term measurements the i...

Страница 16: ...1000Hz sound calibrator in its nominal 94 dB level setting 8 Adjust the CAL potentiometer of the instrument until the display reading for diffuse field is the same as the certified pressure level of t...

Страница 17: ...during measurement the Ov or Un indication is shown on the display to indicate that over range or under range data were included in the sound level measurement values for processing 8 Press the MAX bu...

Страница 18: ...d time weighting settings of the instrument Output voltage 10mV 0 1mV dB Output impedance approx 5k Load impedance 1M The output voltage when the instrument is reading 94dB is nominally 0 94V DC 12 AD...

Страница 19: ...TECH 52N meter fitted with an MC 21 microphone as per IEC 61672 1 and IEC 60651 relative to 1000Hz 15 TYPICAL FREQUENCY RESPONSE OF 65mm WINDSCREEN AT 0 INCIDENCE Freq units 31 5 Hz 63 Hz 80 Hz 100 Hz...

Страница 20: ...nt is a pressure sensitive type it should be equally sensitive in all directions However refraction and cavity effects cause certain microphone directional characteristics at high frequencies The diag...

Страница 21: ...0o 180o 270o 330o 60o 210o 120o 240o 300o Directional characteristics for frequency equal to 2000Hz 25 0 20 0 15 0 10 0 5 0 0 0 5 0 10 0 dB 30 o 0o 90o 150o 180o 270o 330 o 60 o 210o 120 o 240 o 300 o...

Страница 22: ...120o 240o 300o Directional characteristics for frequency equal to 8000Hz 17 APPENDIX A FREQUENCY WEIGHTING NETWORK The SLM 52N provides frequency weightings A and C The electrical characteristics of t...

Страница 23: ...the above mathematical operation looks as follows ei Power of 2 ei2 1 0 T T z ei2 ei2 E rms Input Voltage Output Voltage During sound level measurements the level often fluctuates drastically which wo...

Страница 24: ...e constant which equals CR The response of the index averaging circuit to a single burst signal is shown below C R Input voltage C R Output voltage Equivalent electrical circuit ei ei eo eo ei Input v...

Страница 25: ...han 10dB the values shown in the table below may be used for compensation to estimate the level of the target sound Background noise compensation Display reading difference with and without target sou...

Отзывы: