EN-25
Overload protection :
1000Vrms/600Arms
Accuracy defined for :
Sine wave, ACV>5Vrms / ACA
≧
5Arms, Freq.50~400Hz.
- Only suitable for repetitive signals.
Frequency : Hz
Range
Accuracy
20.0 ~ 399.9 Hz
400 ~ 4000 Hz
±(
0.1% + 5 dgt)
Overload protection :
1000Vrms/600Arms
Sensitivity
:
5Vrms for ACV, 5Arms for ACA ( >400Hz Unspecified)
- The reading will be 0.0 for signals below 10.0 Hz.
Function
Range
Accuracy
ACV / ACA
100.0%
±(
3% + 10 dgt)
Overload protection :
1000Vrms/600Arms
- If ACV<10Vrms or ACA <10Arms, it will display “
rdy
”.
- If the fundamental frequency is out of range 45 ~ 65Hz, the
display will show “
out.F
”.
Inrush Current :
Function
Range
Accuracy
ACA
10.0 ~59.9A
60.0 ~ 600.0A
2.5%
±
2A
2.5%
±
5d
Overload protection :
600Arms
Accuracy defined for :
Sine wave, ACA
≧
10Arms, Freq. 50/60Hz
- Integration time about 100m sec
Active Power : Watts
Function
Range
Accuracy
W~
4.000 kW
40.00 kW
400.0 kW
600 kW
Add the errors of Voltage
and current.
Overload protection :
1000Vrms/600Arms
Accuracy defined for :
Sine wave, ACV
≧
10Vrms, ACA
≧
5Arms
Freq. 45~65Hz, PF=1.00