Used materials
The medical device is manufactured as a welded tubular steel construction. The surfaces are powder
coated or galvanized. All wooden parts are either laminated or lacquered. The surfaces of this
product are unthinkable for the skin from the point of view of health.
10
Service and care
All household cleaners without ammonia and scouring agents are permissible for cleaning the tube
parts, the lying surface and the wooden parts with a damp cloth. Solvents (e.g. nitro) destroy the
coating of the tubes and the lying surface!
Mechanical cleaning (e.g. scraping, sanding) or jet cleaning of the bed is not permitted. All pivots of
the moving parts, including the bearing eyes on the adjustment device, are provided with
maintenance-free slide bearings and must not be oiled or greased.
11
Service life of the product
At an expected average level of use in home care, the service life of the bed is approximately 10 years.
Lack of maintenance and excessive stress on the product can significantly reduce the service life of the
bed. The expected service life in professional nursing home use is approx. 7 years.
12
Disinfection
•
In order to ensure that the bed functions properly, each ISKO bed should be cleaned,
disinfected and checked after each use so that it can be used again immediately.
•
Improper cleaning/disinfection of the bed can cause hazards.
•
Depending on the degree of soiling, we recommend cleaning the bed with a damp cloth or
similar.
•
For wipe and spray disinfection, disinfectants in their intended concentration can be used. (see
manufacturer's instructions)
•
The dilution ratio recommended by the manufacturers in the respective instructions for use
must be used.
Solvents are not permitted.
Abrasives or scouring sponges must not be used.
12.1
Specifications of detergents and disinfectants
•
The working solutions should normally be used freshly prepared.
•
The concentrations given should not be exceeded or fallen below.
•
They must not contain corrosive or caustic components.
•
They must not contain any substances that alter the surface structure or the adhesion
properties of the materials.
•
Lubricants must not be attacked by cleaning and disinfecting agents.
Under no circumstances should soap or washing-active substances be added to the
disinfectant. In the case of products containing alcohol, there is a risk of explosion and fire
when applied over large areas.
The use of unsuitable detergents and disinfectants can cause damage to the surface coating
for which ISKO KOCH GmbH cannot be held liable.