![iSi ISI-799 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/isi/isi-799/isi-799_service-manual_2092808008.webp)
2.
MIN (Minimum)
Specifies the minimum weigh value tracked by analog output.
3.
MAX (Maximum)
Specifies the maximum weigh value tracked by analog output.
4.
TWZERO (Tweak zero)
Adjust the analog output zero calibration. Use a multimeter to monitor the analog
output value, 0VDC for 0-10VDC, 4mADC for 4-20mADC. Edit the number,
press “ENTER” to save new value.
5.
TWSPAN (Tweak span)
Adjust the analog output span calibration. Use a multimeter to monitor analog
output value, 10VDC for 0-10VDC, 20mADC for 4-20mADC. Edit the number,
press “ENTER” to save new value.
1.3.8 Digital Input Output Menu
CONFIG CALIB SERIAL PFORMT ALGOUT DIG IO TEST VERS
DIG IN
DIGOUT
ZERO
0
TARE
1
UNITS
Parameter Descriptions:
1.
DIG IN (Digital Input)
Selects the keypad functions for digital input.
2.
DIGOUT (Digital output)
If select
0
, DO1 = under weight, DO2 = accept weight, DO3 = over weight
If select
1
, DO1 = under and over weight, DO2 = accept, DO3 = over weight