6. SCALE SETTINGS

6.1

PRE-TARE :

With an empty pan Press Zero + 

TARE - “P-tArE” will flash on the 

display. Now press the Zero key 

once and the display shows all 

“000.000”.  Now  Enter  a  value 

(ex.200g),  using  Tare  &  Mode 

keys.  Exit  the  setup  menu  by 

pressing the Zero/Enter key until 

the  scale  comes  in  weighing 

mode.  The  Pre-set  Tare  will  be 

displayed as a negative value (no 

weight  on  the  pan).  Note:  To 

clear the Pre-set Tare value, press 

TARE with the pan empty it will remove it temporarily and to remove the 

Pre Tare value permanently from the memory you have to again go in to 

the settings and make the pre-tare value “000.000” using Tare & Mode 

keys.

6.2

AUTO ZERO :

Press  Zero+TARE  button  The 

display will show A0 – X (0, 1, 2, 

3).This is the auto zero setting. 

You  can  select  the  value  using 

the TARE button. 3 mean 3d in d 

mode  and  3e  in  e  mode 

(remember d =e/10 and e is the 

scale interval. That is e = 2gms and d = 0.2gms) similarly for 3, in d=e/10 

mode it will auto zero upto 0.6gms, and in e mode it will auto zero upto 06 

gms. Default setting of Auto-Zero mode is 3d.

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

6

7

6.3 LED GLOW :

The  display  GLOW  strength  can 

be  increased  or  decreased 

depending  upon  the  user 

requirement.  Increasing  the 

GLOW  of  the  display  will  effect 

the  Battery  back  up  time,  and 

the  battery  life.  The  GLOW 

setting  is  set  to  minimum  by 

Default.  If  the  user  wants  to 

increase  the  GLOW  of  the 

display, it can be done as follows:

When the scale is in the weighing 

mode, While pressing the ZERO 

button, press the TARE button. The display will show “P-TARE” Now press 

the Zero button three times so that the display shows “GLOW 0”. 0 means 

the GLOW mode is currently set to Low. Now press the TARE button, and 

you will notice the GLOW of the display is increased to 1 and 2 level. After 

selecting the level press the ZERO button, until the display comes back to 

the weighing mode.

6.4 SLEEP MODE 

The scale has a power saving mode 

called  the  SLEEP  MODE.  In  this 

mode,  if  the  scale  is  idle  for  one 

minute, that is there is no weighing 

being done, the display will simply 

show the decimal point of the right 

most  digit  on  the  display. 

Whenever any weight is kept, the 

scale will automatically come out of the SLEEP MODE and start displaying 

the weight. 

 

 

Содержание ITA SERIES

Страница 1: ...Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA ITB AS AP AP SERIES Electronic Table Top Weighing Balance ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance TM 14 ...

Страница 2: ... and purposeful destruction are NOT covered by this warranty Technical Service MUST be performed by authorised staff or factory authorized personnel Manufacturer s obligation is limited to replacement of parts that have been so returned and are disclosed to Manufacturer s satisfaction to be defective The provisions of these warranty clauses are lieu of all other warranties expressed or implied and...

Страница 3: ...ernal weight Weight 4 kg Approx Pan Size 230mm x197mm Front Display Type 0 8 Bright red LED display Rear Display Optional Available with 0 56 LED Display Communication Stabilization 3 seconds Power 9V AC Adaptor or Power Cord built in battery for backup to 30hrs 40 hrs with sleep mode active Operating Temp 50 to 98 F 10 to 35 C Dimensions 245 x 210 x 140 w x d x h SOLUTION ERROR DISPLAY ERROR 8 ER...

Страница 4: ...rectly when the bubble scale is in the center of the circle The unit comes with 12V AC adaptor or 220V AC as per model specifications country of use The scale is fitted with a built in 6V 4Ah battery which provides backup of more than 35 50 hours depending on glow setting and sleep mode setting Put a known weight on the machine and make sure that the scale reads the correct weight on the display I...

Страница 5: ...ATION Internal calibration also referred as service calibration requires opening the machine and changing the jumpers It is advisable to do this step only if you have required assistance or the service personnel assisting you to do this Open the battery cover from the bottom of the machine and remove the cover and change the calibration jumper marked as CAL to ON Make sure there is nothing kept on...

Страница 6: ...e changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scale can be calibrated using a known external weight 4 9 7 1 EXTERNAL USER CALIBRATION To start calibration turn the scale off and then turn it on During the count down press and hold down the Zero button do not release until the display shows CAL Make sure there is no weight ...

Страница 7: ...ounting function to come in to the weighing mode It is important to know that SLEEP mode increases the battery back up time by double The Normal backup time when the sleep mode is switched off is 30 Hours and with the SLEEP mode switched on the battery back up time increases to more than 60 Hrs If the USER wants to switch off the SLEEP MODE for some special usage or function it can be done as foll...

Страница 8: ...s 3d ITA Series Table Top Precision Weighing Balance ITA Series Table Top Precision Weighing Balance 6 7 6 3 LED GLOW The display GLOW strength can be increased or decreased depending upon the user requirement Increasing the GLOW of the display will effect the Battery back up time and the battery life The GLOW setting is set to minimum by Default If the user wants to increase the GLOW of the displ...

Отзывы: