background image

SOLUTION

REMEDY

ERROR DISPLAY

SYMPTOM

ERROR

POSSIBLE CAUSE

7. ERROR CODES

8. TROUBLE SHOOTING

·

 

Zero not in range

·

 

Weight not enough for 

the selected setting

·

 

RAM ERROR

·

 

Internal  /  External 

c a l i b r a t i o n   v a r i a n c e  
larger than 2%

·

 

LOW  Battery.  Machine 

displays  message  and 
turns off

 

 

 

 

 

Remove  weight  from  pan 
and restart the machine. If 
the  problem  persists,  the 
machine  may  have  been 
overloaded.

Keep  weight  or  Add  more 
weight  on  pan  for  correct 
results.

S w i t c h   O f f / O n   t h e  
m a c h i n e .   I f   p r o b l e m  
persists call service center.

You  are  trying  to  calibrate 
the scale with an incorrect 
weight.

Battery  is  low.  Connect 
scale to power so that the 
battery can be charged.

Scale will not power 
on

Battery  does  not 
g i v e   e n o u g h  
backup

Scale  does  not 
s h o w   w e i g h t  
accurately

Weight  reading  on 
s c a l e   d o e s   n o t  
stabilize

The  battery  is  defective.  The  battery 
may have past its useful life. (Generally 
6months – 1 year)

If  running  on  batteries,  the  battery 
may have fully discharged. AC Power 
cord not connected. 

Improper calibration.

External  Rubbing  /  Friction.  A  scale 
cannot measure accurately if an object 
is rubbing or pressing against the scale 
platfor m.  Mobile  inter ference, 
Vibration  or  wind  could  be  other 
possible reasons

Connect  scale  to 
power source. 

Recharge  batter y 
Change the battery.

Move the scale away 
from  mobile  signal, 
breeze  or  vibration 
areas. 

Calibrate  the  scale 
again  to  and  check 
f o r   t h e   p r o b l e m  
again.

2 INSTALLATION

2.1 Unpacking

Save the packaging material. This packaging ensures the best possible 
protection for the storage or transport of the product.

2.2 Placing your scale

2.3

Leveling the scale

2.4

Power Source

2.5

Calibration Check

Unpack and verify that the following components have been 
included:
a) ITA  Series (G) Precision Balance
b) Instruction Manual
c) AC Adapter 

Before you install the scale, identify the best location for the equipment. 
The proper environment enhances its operation and longevity. Keep in 
mind the following factors, which might have a negative influence on 
the scale's operation : 
a) Vibration  :  Vibration  diminishes  the  scale's  ability  to  measure 
accurately. Electrical machinery such as conveyors and drill presses can 
cause inaccurate and non-repeatable readings. The scale may also read 
inaccurately if it is not leveled properly. The scale incorporates a Special 
weighing mode where the scale can reject readings with vibration up to 
some extent.
b) Drafts / Air currents : Moving air can cause the scale to read wind 
movement  as  an  additional  force  and  cause  inconsistency  in  the 
weighing results.
c) External Rubbing / Friction: A scale cannot measure accurately if an 
object is rubbing or pressing against the scale platform.
d) Moving the Scale: If you move the location of the scale it may require 
recalibration. Please check the calibration when the scale is moved to a 
new location.

Level the scale by turning the adjustable feet. It is leveled correctly 
when the bubble indicator is in the center of the circle.

The unit works through 9V AC Adaptor. The scale is fitted with a built 
in 6V 4Ah battery which provides backup of more than 45 ~ 50 hours 
depending on glow setting and sleep mode setting.

Put a known weight on the machine and make sure that the scale 
reads the correct weight on the display. In case of a variation in the 
display reading, it is recommended to calibrate the scale again with 
the procedure explained in section 6 later in the manual.

ITA Series – Table Top Precision Weighing Balance

ITA Series – Table Top Precision Weighing Balance

2

11

Содержание ITA-310-G

Страница 1: ...Scale Users Guide ISHTAA SCALES INC INDIA www ishtaascales com ITA Series Table Top Precision Weighing Balance ITA Series G Precision Table Top Weighing Balance TM ...

Страница 2: ...nce ITA Series Table Top Precision Weighing Balance Contents 1 Technical Data 1 2 Installation 2 3 Controls Functions 3 4 User Operation 4 5 Scale Settings 5 6 Calibration 7 7 Error Codes 11 8 Trouble Shooting 11 9 Warranty 12 GS ISI 13 04 09 1T ...

Страница 3: ...CATE Customer Name ______________________________________________ _____________________________________________________________ _____________________________________________________________ Phone _______________________ Mobile _________________________ Date of Purchase ______________ Invoice DC No __________________ Model ______________________ Sl No __________________________ Dealers Agent Name _...

Страница 4: ... Unpack and verify that the following components have been included a ITA Series G Precision Balance b Instruction Manual c AC Adapter Before you install the scale identify the best location for the equipment The proper environment enhances its operation and longevity Keep in mind the following factors which might have a negative influence on the scale s operation a Vibration Vibration diminishes ...

Страница 5: ... selected setting and exit from setting mode External Calibration of Scale STEP DISPLAY PROCEDURE 1 2 3 4 5 6 Press and hold the tare button while switching on the scale You will see CAL 0 on the display for a second and then some counts will be displayed Press the mode button once you will see Load on the display Keep a known weight on the pan for eg 200 gm and press mode button The display will ...

Страница 6: ...SCALE ON OFF The scale is switched ON OFF by using the Power switch located on the right side of the scale Note Make sure the weighing pan is empty before turning on the scale 4 2 TARE ZERO The machine has a common Tare Zero button TARE ZERO BUTTON This button is used for taring a weight to display 0 000 Place a container on the weighing pan say 200g then press the TARE button The weight is zeroed...

Страница 7: ...ight most digit of the display and if the DSP 0 setting is set to NO the display will show all the 4 zeros on the display It is important to know that DSP 0 mode if set to YES increases the battery back up time STEP DISPLAY PROCEDURE 4 5 6 7 8 9 You will now see P off on the scale Press Mode button and you can select the P OFF option described in section 5 3 as YES NO Default option is Yes You can...

Страница 8: ...r the scale may require re calibration Temperature changes geographic gravity variations altitude changes and abuse are few reasons why a scale may need recalibration The scales can be calibrated using a known external weight Switch off the scale Press the MODE button and switch on the scale You will see PASS Press MODE again and you will see 0000 on the display The password is 1072 and is to be e...

Отзывы: