3.2.1 Main screen (Channel List)
After start up of the NHS the display will show a Main screen similar to the following image which consists of 5 elements:
1. Left function button: main menu
Shows the function of the left button. If the user pushes the left button in Main screen, the display will show the Menu
screen.
2. Right function button: display mode
If the user pushes the right button in Main screen, the Main screen will switch to the next display mode, which is shown
directly above the channel list.
Following display modes are available:
1. VI-MEAS
➜
voltage and current – measured values
2. VI-SET
➜
voltage and current – set values
3. V-MS
➜
voltage – measured and set values
4. I-MS
➜
current – measured and set values
3. Channel numbers and running states
This is part of the channel list (4) and shows the corresponding channel number. The background color signals the
running state of each channel. The background colours means the following:
Black
➜
the channel is switched off
Green
➜
the channel is switched on
Yellow
➜
the channel is switched on but no regulation (CV, CC) is active
Orange
➜
the channel is switched on but has unmasked error event(s)
Red
➜
the channel has at least one masked error event
4. Rotary encoder button function
If the user presses the button of the rotary encoder, the menu will switch to the channel menu of the marked channel.
1
Left function button: main
menu
2
Right function button:
display mode
3
Channel numbers and
running states
4
Channel list and rotary
encoder function
5
Rotary encoder button function
NHS – NIM standard High Voltage Power Supply | Last changed on: 16.01.2020 |
www.iseg-hv.com
13/30