2.3 Options
OPTIONS
OPTION CODE
EXAMPLE
ITEM CODE HEX CODING
POLARITY
Positive:
x = p
,
Negative
x = n
EHS 82 05
p
VERY LOW NOISE (only EHS CG Series)
VLN
010
SINGLE CHANNEL INHIBIT - down
ID
400
SINGLE CHANNEL INHIBIT - up
IU
800
NEGATED LOGIC INHIBIT ID, IU
N
80
VOLTAGE CORRECTION by TEMPERATURE
VCT
008
LOWER TEMPERATURE COEFFICIENT
TC
T10
004
ACTIVE SAFETY LOOP
SLA
001
INTERNALLY POWERED SAFETY LOOP
SLP
002
ONLY ONE CURRENT RAGE FOR HIGH
PRECISION MODULES
1CR
020
200 V ISOLATION FOR FLOATING GND
F02
100
2,000 V ISOLATION FOR FLOATING GND
F20
200
LOWER OUTPUT CURRENT
(1
L
(I
nom
= 100 μA)
–
Notes:
1)
Requires option „1CR“.
Table 9: Technical data: Options and order information
3 Handling
3.1 Connection
The supply voltages and the CAN interface are connected to the module via a 96-pin connector on the rear side of the module.
The physical address of the module, determined by the slot position in the crate, is also accessible via this connector
Modules and crate controllers with different settings of bit rate do not work on the same CAN-Line.
INFORMATION
INFORMATION
Note: For proper operation the module must be configured with the correct CAN bitrate, which meets the
configuration of the crate controller, the module will be used with. The delivery condition is shown on the
modules typeplate (side plate of the module).
Typically newer iseg crate controllers (CC24, CC23, CC238) are delivered with 250kBits/s standard. Wiener
M-POD Controller and older iseg hardware is set on 125 kBit/s standard bitrate.
INFORMATION
INFORMATION
Note: EHS modules with Common Floating Ground (CFG) will be delivered with a jumper, which connects the
module-GND with the crate-GND. To operate in CFG configuration the jumper (CG-CFG) on the module back
must be removed. (
)