7.1. Status information
xxx:
ON<SP>
Output voltage according to set voltage
OFF
Channel is switched off via the front panel switch
MAN
Channel is on, set to manual mode
ERR
V
max
or I
max
is / was exceeded
INH
Inhibit signal was / is active
QUA
Quality of output voltage not given at present
L2H
Output voltage increasing
H2L
Output voltage falling
LAS
Look at Status (only after G-command)
TRP
Current trip was active
If output voltage has been shut off permanently (by ERR or INH at ENABLE KILL or TRP) the command ”
READ STATUS WORD
”
must be executed before the output voltage can be restored.
7.2. Error codes
????
Syntax error
?WCN
Wrong channel number
?TOT
Timeout error (with following reinitialization)
?<SP>UMAX=nnnn
Set voltage exceeds voltage limit
Notes:
n = Non-negatives Integers
EHQ High Voltage Power Supply series | Last changed on: 11.05.2020 |
www.iseg-hv.com