19
USE AND MAINTENANCE MANUAL
MILLENNIUM MY2019
EN
428001137037
9.
MAINTENANCE
The
Staff in charge of the appliance
must control and respect the expiry dates for maintenance, given in the table
below, contacting the authorised
Technical After-sales assistance
when indicated.
OPERATION
FREQUENCY
ORDINAR
Y
EXTRAORDINAR
Y
PERSONNEL
AUTHORISED
D
epending o
n U
se
an
d Necessit
y
Monthly
Ev
er
y 6
m
onths
An
nual
CLEANING THE EXTERNAL SURFACES
X
X
USER
CLEANING THE ACCESSIBLE INTERNAL PARTS
(without the use of tools)
X
X
CONTROL POWER SUPPLY CABLE,
PLUGS AND / OR ELECTRICAL SOCKETS
X
X
CONTROL INTEGRITY OF SEAL
X
X
CONDENSING UNIT FILTER CLEANING
(whenever present)
X
X
DEFROSTING WATER COLLECTION TRAY CLEANING
X
X
TECHNICAL
ASSISTANCE
CONDENSER CLEANING
X
X
X
CHECK COMPRESSOR OIL LEVEL
(whenever present)
X
AIR TANK DRAINING
(whenever present)
X
X
CONTROL PNEUMATIC CONNECTIONS
(whenever present)
X
X
CONTROL INTEGRITY PIPE COOLING SYSTEM
X
X
INSPECTION OF CABLES
INTERNAL CONNECTIONS AND POWER
X
X
CONDENSATE DRYING SPONGES CLEANING
(whenever present)
X
X
LAMP / LED REPLACEMENT
(whenever present)
X
REPLACEMENT OF CONTROL PANEL
(electronic control unit - thermostat - etc.)
X
REPLACEMENT OF POWER SUPPLY CABLE, PLUGS AND/OR ELECTRICAL SOCKETS
X
ATTENTION
After all maintenance it is
mandatory
to perform all electrical safety tests in accordance with
the IEC EN 50106 Standard.