Quick User Guide – IRIScan
TM
Executive 4
5. Maintenance
5.1 Calibrate the scanner
Calibrate only if the scanned image looks fuzzy, the color looks abnormal (e.g., darker than normal), after
more than 8,000 scans, or the scanner has been stored without use for a long time. The calibration process
resets the Contact Image Sensor to resume its color and sharpness precision.
1.
Connect the scanner to your computer.
2.
Start the
Capture Tool
.
Click
Start
>
All Programs
>
IRIS
IRIScan Executive 4 Scanner > Capture Tool
Select your driver type and scanner model and click
OK
.
3.
The Capture Tool dialog box will be displayed.
4.
Click the “
Setup
” button to open the Scanner Properties dialog box.
5.
Click the “
Settings
” tab and then click the “
Calibration
” button.
6.
The “
Insert calibration sheet
” dialog box appears.
7.
Hold the edges of the calibration sheet
(included)
and insert it
with the text
facing up
and the white
area feeding first into the feeder (As shown below).
8.
Click “
OK
” in the “
Insert calibration sheet
” dialog box to start calibrating the scanner.
WARNING:
If the calibration failed, the led will be flashing and an error message will be displayed on
the screen. Disconnect and reconnect the scanner and then repeat the calibration.