Section 2
Getting Started
2.11
2.6.2 Cabling Length
The IEEE 488 specification allows for cabling distance between bus
devices of 2 meters (approx. 7 feet). The total worst case distance from the
two furthest devices is not allowed to exceed 20 meters (approx. 70 feet).
The addition of the Isolator488 does not increase this distance.
2.6.3 Signal Line Direction
The signal lines and their direction include....
Data Lines:
DIO1 - DIO7
Bidirectional
Handshake Lines:
DAV
Bidirectional
NRFD
Bidirectional
NDAC
Bidirectional
Command Lines:
SRQ
To Controller Bus
ATN
To Instrument Bus
EOI
Bidirectional
REN
To Instrument Bus
IFC
To Instrument Bus
The majority of the command lines have their signal direction fixed.
This forces bus control from the CONTROLLER bus.
When Attention (ATN) is asserted, the data lines, talker handshake
line (DAV) and the EOI line direction is forced from the CONTROLLER
to the INSTRUMENT port. The listener handshake lines (NRFD and
NDAC) are forced from the INSTRUMENT to the CONTROLLER port.
The commands are sent through the Isolator488 to the INSTRUMENT
port.