Section 2
Getting Started
2.8
WARNING
Never open the Isolator488 case while it
is connected to the AC line. Internal
voltage potentials exist which could
cause personal injury or death.
1.
Place the Isolator488 on a flat surface. Remove the six screws on top of the
case and remove the top cover. Located next to the rear panel power connector is
switch S2 and fuse F1.
2.
Insert the tip of a small screwdriver into the slot of the switch and move the
switch to the left or right so that the desired line voltage selection appears on the
switch. Make note of the new setting for later reference.
CAUTION
Use the 1/2 amp fuse, FU-1-.5, for 115
volt operation. Use the 1/4 amp fuse,
FU-1-.25, for 230 volt operation. Failure
to use the correct fuse could result in
damage to the Isolator488.
3.
Replace the 1/2 amp fuse, FU-1-.5 (for 115 volt operation), with the 1/4 Amp
fuse (for 230 volt operation), part number FU-1-.25, that is included in the
Isolator488 Accessory Kit.
4.
Carefully reassemble the unit.
WARNING
The Isolator488 is intended for INDOOR
USE ONLY. Failure to observe this
warning could result in equipment
failure, personal injury or death.