Technical Specifications
Minimum Resolution
0.1 ppm
Detectable Range
0.1 ppm to 5,000 ppm
Response Time
T90 < 2 seconds
Accuracy
± 5% display reading ± one-digit (at calibration point)
Intrinsically Safe Approvals
•
II 1G Ex ia IIC T4 Ga
•
Tamb = - 15
o
C ≤Ta ≤+45
o
C (with Lithium ion Battery Pack)
•
Tamb = - 15
o
C ≤Ta ≤+45
o
C (with Alkaline Battery Pack)
•
IECEx ITS 22.0025X ITS-I22ATEX35111X
•
ITS22UKEX0635X
•
3193491 conforms to UL Std. 913, 61010-1 &
•
Certified to CAN/CSA Std. C22.2 No. 61010-1
Battery Life
Typical operating time of 24 hours
Typical charge time of 8 hours
Alkaline 3 x AA: Typical battery life of 8.5 hours
Lamps
10.6eV krypton (standard)
Data Log
Including date / time: 80,000
Communication
Direct USB 1.1
Calibration
2- and 3-point calibration (via calibration kit accessory)
Alarms
Flashing LED’s
amber (low alarm) red (high alarm)
Sounder 95 dBA at 300 mm (12”)
Vibration on alarm
Flow Rate
Temperature
Weight & Dimensions:
≥ 220 ml/min
Operating:
-20 to 60
o
C / -4 to 140
o
F (non -intrinsically safe
environment)
Certified to:
15 to +45
o
C / +5 to +113
o
F (Intrinsically Safe
environment)
Storage:
-25 to 60
o
C / -5 to 140
o
F
Humidity: 0-99% RH (non-condensing)
Instrument Weight:
870g
Instrument Dimensions:
370 mm (H) x 91 mm (W) x 60 mm (D)
Содержание TIGER XTL
Страница 3: ...TIGER XTL Instrument User Manual V1 0 Pioneering Gas Sensing Technology Page 3 of 56 ionscience com...
Страница 4: ...TIGER XTL Instrument User Manual V1 0 Pioneering Gas Sensing Technology Page 4 of 56 ionscience com...
Страница 11: ...TIGER XTL Instrument User Manual V1 0 Pioneering Gas Sensing Technology Page 11 of 56 ionscience com...