TIGER XTL Instrument User Manual V1.0
P
ion
eering Gas Sensing Technology.
Page 40 of 56
ionscience.com
Custom Calibration Procedure
From the main running screen, Press the
Options
soft key
on your TIGER XTL to access the
adjustable features.
Then use the
Up
or
Down
key to select calibration.
Press
Enter
to confirm selection.
Select
Custom Calibration
and press
Enter
to confirm.
On confirming the selection, the user is presented with a 30 second count down.
Remove both caps from the Carbon Filter (A-31057) and then fit onto the probe of your TIGER XTL.
Press
Enter
to start the ‘zero’ countdown. At the end of the count down a tick '
✓
' will appear,
indicating that the Zero has been accepted. Disconnect the Carbon Filter Assembly and replace the
cap ends. The useful life of the Carbon Filter Assembly will be shortened if the Carbon Filter
Assembly is open to atmosphere for prolonged periods.
Press
Enter
again and the gas and concentration for Span 1 (previously set up in TIGER LT PC) are
displayed along with a 30 second countdown.
Attach the ‘Span 1’ gas using the Calibration
Adaptor (861476) (see ’Accessories’) and press
Enter
to start the Span 1 countdown. At the end of
the count down a tick '
✓
' will appear, indicating that the Span 1 has been accepted. Press
Enter
and
the calibration is complete.