![Invitrogen Qubit 2.0 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/invitrogen/qubit-2-0/qubit-2-0_user-manual_2085418008.webp)
8
Product Specifications
Qubit
®
2.0
Fluorometer
Specifications
Environmental Conditions
Operating power:
100–240 VAC, 0.4 A
Frequency: 50–60
Hz
Electrical output:
9 VDC, 1.33 A
Installation site:
Indoor use only
Operating temperature:
15–40°C
Maximum relative humidity: 20–80%,
non-condensing
Instrument Specifications
Instrument type:
Benchtop fluorometer
Instrument dimensions:
5.3” (w) × 8.5” (l) × 1.6” (h)
(13.3 cm × 21.3 cm × 4 cm);
oval
shaped
Weight:
11.5 oz (340 g)
Dynamic range:
5 orders of magnitude
Processing time:
≤ 5 seconds/sample
Light sources:
Blue LED (max ~470 nm)
Red LED (max ~635 nm)
Excitation filters:
Blue 430–495 nm
Red 600–645 nm
Emission filters:
Green 510–580 nm
Red 665–720 nm
Detectors:
Photodiodes: measurement
capability from 300–1,000 nm
Calibration type:
2- or 3-point standard
Tube type:
0.5 mL Real Time PCR
(polypropylene)
tubes
Warm-up time:
2 seconds
Qubit
®
2.0 USB Drive:
2 Gigabyte