Calibrating the Qubit
®
2.0 Fluorometer,
Continued
Running New
Standards for
Calibration,
continued
4.
Insert Standard #1 into the Sample Chamber and press
Read
.
Ensure that you are using the Standard #1 appropriate for the
assay you are performing. The reading takes approximately
3 seconds.
5.
Insert Standard #2, and press
Read
. Ensure that you are using
the Standard #2 appropriate for the assay you are performing.
6.
If you are performing the Qubit
™
protein assay, which
requires a three-point calibration, insert Standard #3 when
prompted, and then press
Read
.
The calibration is complete after Standard #2 is read (or after
Standard #3 is read for the Qubit
™
protein assay). The new
standards graph with data points for standards connected by
a line appears on the screen.