9.6.3
Guidance and manufacturer’s declaration—electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The user must
ensure that the device is used in such environment.
Immunity test
IEC 60601
Test level
Compliance level
Electromagnetic
environment
—
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%. Try to use the
device in the environment
specified. If the condition
of the environment cannot
be determined, other
preventive measures must
be taken, such as using an
antistatic device,
discharging static electricity
from the body or wearing
antistatic clothing.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV
1 kV for input/output
2 kV
1 kV for input/output
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV differential
mode
2 kV common mode
1 kV differential
mode
2 kV common mode
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % UT
(>95% dip in UT)
for 0.5 cycle
40 % UT
(60% dip in UT)
for 5 cycles
70 % UT
(30% dip in UT)
for 25 cycles
<5 % UT
(>95% dip in UT)
<5 % UT
(>95% dip in UT)
for 0.5 cycle
40 % UT
(60% dip in UT)
for 5 cycles
70 % UT
(30% dip in UT)
for 25 cycles
<5 % UT
(>95% dip in UT)
Mains power quality should
be that of a typical
commercial or hospital
environment.
If the user of the device
requires continued
operation during power
interruption, it is
recommended that the
device is powered by an
uninterruptible power
source.