33
EN
CLEANING AND MAINTENANCE
CLEANING OF THE MAIN UNIT
Use a clean dry cloth and mild detergents for cleaning the unit.
ATTENTION: during cleaning operations, make sure the inter-
nal parts of the unit are not in contact with liquids and the power
cord is not connected.
CLEANING AND DISINFECTION OF THE ACCESSORIES
After each treatment:
The nebulizer, mouthpiece and nosepiece (optional) must be disinfected by boiling
with water (max. 10 minutes) or in autoclave (max. 121°C). We suggest you disin-
fect the mask (optional) using cold disinfecting liquids (solutions with hypochlorite).
Dry the nebulizer using a clean cloth; after connect it to the air outlet (pic. A/4),
switch the device on and let it work freely for 10-15 minutes.
Do not boil nor autoclave the air tube and the mask (optio-
nal).
Following carefully the cleaning and disinfecting instructions is
very important for the product performance and for a successful
therapy.
REPLACEMENT OF THE NEBULIZER
The nebulizer must be replaced after a long period of inactivity, or when it is
obstructed by dry medication, dust, etc.
Your nebulizer should last from 6 months to 1 year with general use.
Use original nebulizers only.
REPLACEMENT OF THE FILTER
The filter must be replaced approximately after 30 hours of use, or when it turns
dirty; open the filter holder (pic. A/5) to replace the filter (pic. A/8) with a new
one. Use original filters only.
Do not use the device without filter.
man STARTOS PRO EN 28/37 16-07-2008 15:32 Pagina 33