9 Troubleshooting
9.1 Identifying and repairing defects
There are audio and visual alarms present on the control unit.
Problem / Alarm
Cause
Solution
Mattress CPR hose disconnected.
Connect CPR hose connector, lock it in place.
Mattress not inflating
(not alternating
properly).
Has power and fuse is good, control unit does
not come on.
Send control unit back to Invacare for repair.
Major leak in air cell.
Replace leaking air cell.
CPR hose or tube connectors kinked or split.
Unkink or replace split CPR hose or tube
connectors.
Not alternating, rotor failure.
Send control unit back to Invacare for repair.
No air (control unit failure).
Send control unit back to Invacare for repair.
No power.
Control unit off.
Check power source, turn unit on.
Power cord disconnected.
Connect power cord and ensure the power source
is on.
No power in the power source.
Check the power source has power and turn it on.
Power outage.
Wait till the power source has power.
Fuse blown.
Change fuse on power inlet connector with spare
fuse or identical replacement only (consult a
trained engineer if you are unsure how to change
a fuse).