IVCNORBE221119
12
Invacare® Soft Tilt
Status indicators
Active Program
Display
LED
Note
AUTO
AU
AUTO LED
active
Program 1
P1
P1 LED active
Program 2
P2
P2 LED active
Timer
P1/XX and
P2/XX
The display
changes between
the selected
program and the
time to the next
reposition.
-P1/01 = program
1 and 01 minute
before next repo-
sition.
-P2/05= program
2 and 05 minutes
before next repo-
sition.
4.7.1 Auto program sequence
• Starting from the flat position, both wings immediately
begin to raise simultaneously.
• Left wing stops at 12° and the right wing stops at 40°.
• 5 seconds after the right wing has reached 40° the left wing
lowers to 0°.
• This position A is maintained for 30 minutes.
• Simultaneously the left wing begins to raise and the right
wing begins to lower.
• The right wing stops at 12° and the left wing continue to
raise up to 40°.
• 5 seconds after the left wing has reached 40° the right wing
lowers to 0°.
• This position B is kept for 30 minutes.
The cycle will continue shifting between position A and B, until
the program is stopped.
IMPORTANT!
- Before an emergency release of the wing, re-
move the plug from the mains socket.
4.9
Emergency release of the wing
In case of a power or motor failure, an emergency release of the
wing could be necessary.
CAUTION !
Risk of injury
- A minimum of two persons is required for an emergency
release of the wing.
- When releasing the wing it might lower fast. Keep clear
the area under the wing and arms.
1. Both persons hold the elevated wing.
2. One of them locates the motor pin in question and pulls out
the safety pin.
3. After the pin is removed, the motor will slide out of the sup-
port and slowly lower the wing.
The hand control will not work when a program is
active.
2. To stop a running program press the ON/OFF
A
button.
The battery is charging when the system is connected to the
power supply and turned on.
Charging of the battery is indicated by a dot in the bottom right
corner of the display.
The battery will be fully charged after 12 hours.
4.8.
Charging the battery
The battery is NOT charging when error E03 or E04
occurs.
4.7.2 Pause automatic operation
It is possible to pause an automatic operation and use
the hand control as in manual mode.
1. Press and hold for 3 sec.
R
button, while an automatic pro-
gram is running, the wings will move to the flat position. Now
the manual mode can be used.
2. The activated program before the pause action, will proceed
after 30 minutes of inactivity, 10 sec before the program starts
a BEEP is heard, if the inactivity period should be extended
press any button on the hand control after the BEEP.
3. If the automatic program should start again right away, press
and hold
R
for 3 sec to activate.
Содержание Soft Tilt