5 ASSEMBLY
Manual/Electric Portable Patient Lift
18
Assembling the Electric Actuator to the Boom
1.
Remove the shoulder bolt, washer and nut from the mounting
bracket on the boom assembly.
2.
Unpack the pinch guard from the patient lift carton.
3.
Cut the plastic-wrap that secures the boom and mast together.
4.
Lift-up on the boom and place it on your left shoulder.
5.
Let the actuator rest on your right-side of your chest and rotate the
shaft extension of the actuator assembly until it lines-up with the
mounting holes in the boom assembly.
6.
Place the pinch guard over the shaft extension of the actuator.
7.
Align the holes of the boom assembly mounting bracket with those
of the actuator and insert the bolt. Secure with nut.
8.
Plug in the electrical connectors from the electric actuator to the bottom of the battery assembly.
9.
Plug the hand control control into the bottom of the battery assembly.
CAUTION
DO NOT overtighten the nut and bolt. This damages
the mounting bracket.
The bottom of the electric actuator assembly will already
be assembled to the mast mounting bracket.
Pinch Guard
Mounting Bracket
Turn 90°
Be sure that the bolt is completely through the holes of the boom assembly mounting bracket and the actuator assembly. The boom
assembly will pivot easily if the mounting hardware is aligned properly when the boom assembly is secured to the mast.
The bottom of the actuator assembly will already be assembled to the mast mounting bracket.