60126903-A
24
1. Remove the cannula from its packaging.
2. Perform one of the following:
l
For use without a humidifier bottle—Attach the patient’s nasal cannula supply tube
B
to the oxygen
outlet connector
C
on the concentrator.
l
For use with a humidifier bottle—Attach the patient’s nasal cannula supply tube
B
to the humidifier
bottle outlet
D
.
3. After assembly, turn on the concentrator. Refer to
5.4.3 Turning the Concentrator On, page 24
WARNING!
Risk of Injury or Damage
Failure to properly install the humidifier bottle or other accessories to the concentrator will impact the
flow of oxygen.
5.4.3 Turning the Concentrator On
5-30 min
1. Plug in power cord to an electrical outlet.
2. Press the start/stop button
A
on the display to turn the unit on.
The panel lights will come on and the concentrator will beep, confirming that the indicators are
functioning properly. The start/stop button LED on the display will remain on.
The start-up symbol will rotate clockwise and blue light will illuminate when the start/stop button is
pressed. Allow between 5 and 30 minutes for the unit to achieve full oxygen concentration. When the
device is ready, the check mark and green light will illuminate.
If the concentrator is moved into an ambient temperature of 68° F (20° C) after being stored at its
minimum storage temperature of -13° F (-25° C) or at its maximum storage temperature of 158° F (70°
C) it may take up to 20 minutes to begin operating.
Invacare® Platinum® 5NXG Oxygen Concentrator