60126903-A
38
Alarm
Error
Code
Solution
Symbol
Indicator
Light Color
Sound Indication
High Temperature
Alert
L4000
1.
Inspect for blocked vents
2.
Ensure device is in specified
operating environment
3.
If alarm is still present,
contact provider
White
with
yellow
dot
Single audible beep with
less than 20 second
interval.
Alarm
Error
Code
Solution
Symbol
Indicator
Light Color
Sound
Indication
Shut-Down Alarms
Power Loss 3 times in a row
H1000
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Low Oxygen Concentration
Shut Down
H2040
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Low Oxygen Concentration
Shut Down Diagnostics
H204x
1. Check if flowmeter is set
between 1.0-5.0 L/min
2. Inspect cabinet filter
3. Inspect for kinks or blockages
(kinked cannula, blocked vents,
etc.)
4. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Pressure Sensor Out of
Range High
H3000
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Pressure Sensor Out of
Range Low
H3010
1. Restart unit using stop/start
button
2. If alarm is still present, contact
provider
White
with red
dot
Constant
Audible
Alarm
Invacare® Platinum® 5NXG Oxygen Concentrator