SECTION 3—USING THE STANDARD STYLE SLING
10
Attaching the Chains or Straps to the Standard Style Sling
NOTE: For this procedure, refer to FIGURE 3.3.
NOTE: Each sling will work with either a pair of chain assemblies or a pair of strap assemblies.
CAUTION
When using either the chains or straps to connect the sling to the patient lift, the short-
est of the two sections of the chains or straps MUST be attached to the back section of
the sling. Using the long section will leave little or no support for the patient’s back.
1. With the patient laying on his/her back and positioned properly in the sling, lay the chains
or straps on the bed next to the sling. Refer to Detail “A”.
2. The short section of chain or strap will be placed parallel to the patient’s back. This will
position the longer section of chain or strap from the waist to the mid section of the thigh.
3. The “S” hooks on the ends of the chains or straps are inserted into the metal support bar
of the sling.
4. The “S” hook MUST be inserted through the hole in the support bar with the open end
facing away from the patient. Refer to Detail “C”.
5. Position the lift for use. Refer to Attaching the Standard Style Sling to the Patient Lift on
FIGURE 3.3
Attaching the Chains or Straps to the Standard Style Sling
MODEL NUMBERS
DESCRIPTIONS
9070
Strap Kit
9071
Chain Kit
S-Hook Open End Away
from Patient
S-Hook Open End
Away from Patient
O-Ring
Short Section of Chain
S-Hook Open End Away
from Patient
Short Section
Long Section
Long Section of Chain
D-Ring
Adjustable
DETAIL “C”
DETAIL “B”