5.6 Resetting the System
Only for beds with Trendelenburg and / or
Anti-Trendelenburg function
If a height adjustment actuator is newly connected or
reconnected to the control unit, the system needs to be
reset to ensure proper function.
C
B
A
To reset the system with the hand control:
1.
Insert the magnetic key
A
in the slot on the side of
the hand control.
2.
Press and hold button
B
and
C
for 3 seconds to
initiate the system reset and hold the buttons pressed
until the following sequence is complete:
a. The height adjustment actuators move to their
maximum position (bed in highest position).
b. All actuators of the mattress support sections
move to their minimum position (mattress support
completely flat)
c. The height adjustment actuators move to their
minimum position (bed in lowest position).
1657008-A
15