Invacare® MK6i™ Electronics
Sets the tipsy switch angle positions
•
Use Up and Down arrows to move (TILT) to the HOME
position, (or Slow Down, DLO, or Max Back), then press
SELECT. Press right arrow to change actuators.
•
HOME; 0° to XX°, the position within which Elevate UP will
function.
•
SLOW: Position at which driving speed is reduced to 20%
•
DLO: Position for Drive Lock Out
•
MAX BACK ANGLE: Sets the maximum back angle the system
will travel
•
SET DEFAULT: Choose “PRESET” in SET DEFAULT menu to
return to factory default.
•
RESET: Intended for future use. Factory Default Settings are:
– Home 7.5°
– Slow 15°
– *DLO 30°
– Max Back Angle 78° (168°)
*NOTE: 30° DLO (Default) is only for models using the
M610i module for powered seating
Tipsy Operation
•
The seating system must be in the must be in the “HOME”
position (default = 0° — 7.5°) to allow elevate up.
•
The system can elevate down in any seat position.
14.3.3
Drive Lock Out
DRIVE LOCKOUT
DRIVE 1
ON
OFF
Turns “DRIVE LOCK OUT” On or Off in each Drive
•
Select All Drive or an individual drive and choose “ON” or
“OFF”
80
1141471-L-00