SECTION 3—MAINTENANCE AND TROUBLESHOOTING

Part No 1148139

23

MA90 Series

SECTION 3—MAINTENANCE AND 
TROUBLESHOOTING

Cleaning the System

 

WARNING

Before cleaning or disassembling the MA90 series, check the underside of the 

mattress folds for sharp objects such as scissors, needles, etc. These objects should 

be removed and discarded before proceeding with further cleaning or disassembly; 

otherwise injury or damage to the product may occur.
Because of potential risk of infectious exposure, cleaning with the patient on the bed 

is not recommended.
All equipment should be inspected. Any item that is visibly soiled with the patient's 

blood or other body fluids should be properly cleaned or removed.
Laundry workers should treat all soiled bedding as if it were contaminated with 

pathogenic microorganisms.

1. Remove

 

the

 

bedding.

2. If

 

necessary,

 

inflate

 

the

 

mattress.

3. Ensure

 

that

 

the

 

control

 

unit

 

is

 

off.

4. Unplug

 

the

 

power

 

cord

 

from

 

the

 

wall

 

outlet.

5. Ensure

 

that

 

the

 

underside

 

of

 

the

 

mattress

 

is

 

clear

 

of

 

all

 

sharp

 

objects.

6. Examine

 

the

 

surface

 

of

 

the

 

control

 

unit

 

and

 

mattress

 

assembly

 

components

 

for

 

visible

 

blood

 

or

 

body

 

fluids.

 

7. Perform

 

one

 

of

 

the

 

following:

• If

 

blood

 

is

 

present,

 

decontaminate

 

the

 

product.

 

i. Remove

 

all

 

visible

 

soil

 

with

 

disposable

 

paper

 

towels.

ii. Scrub

 

the

 

area

 

with

 

freshly

 

prepared

 

effective

 

phenolic

 

detergent

 

disinfectant

 

solution.

• If

 

blood

 

is

 

not

 

present,

 

remove

 

any

 

soil

 

from

 

the

 

cover

 

with

 

paper

 

towels.

NOTE:

 

If

 

soiled,

 

the

 

cover

 

should

 

be

 

removed,

 

cleaned

 

and

 

decontaminated.

8. Perform

 

the

 

following

 

steps

 

to

 

clean

 

the

 

mattress

 

cover:

A. Using

 

a

 

clean

 

sponge

 

or

 

paper

 

towel,

 

wipe

 

down

 

the

 

cover

 

surface

 

with

 

a

 

dilute

 

detergent

 

solution

 

of

 

quaternary

 

cleaner

 

disinfectant

 

or

 

other

 

germicidal

 

detergent

 

solution.

B. Remove

 

the

 

cover

 

and

 

launder

 

it

 

using

 

the

 

following

 

method:

Содержание microAIR MA90 Series

Страница 1: ...BEFORE using this product read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com MA90 Series MA90Z Alternating...

Страница 2: ...facturers have not been tested by Invacare and are not recommended for use with Invacare products SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which coul...

Страница 3: ...PERATION 10 Installing the Mattress Replacement System 10 Installing the Side Rails 11 Installing the Power Unit 11 Connecting the Hose 12 Connecting the Power Cord 13 Using the Front Panel 14 Power B...

Страница 4: ...g the System 24 Changing the Filter 25 Troubleshooting 25 LIMITED WARRANTY 28 REGISTER YOUR PRODUCT The benefits of registering include 1 Safeguarding your investment 2 Ensuring long term maintenance...

Страница 5: ...low fuses MODE OF OPERATION Continuous MA90Z MA95Z WEIGHT CAPACITY STANDARD MATTRESS BARIATRIC MATTRESS 350 lbs 1000 lbs PRESSURE ZONE 2 MAXIMUM FLOW 50 15 LPM 1275 100 LPM MAXIMUM PRESSURE 35 5 mmHg...

Страница 6: ...L UNIT CONNECTION Quad Flow Four Single Quick Coupling Triple Flow Two and One Single Magnetic Quick Coupling CONTROL UNIT PACKAGING 1 piece box AIR FILTER None Charcoal Air Filter with Fire Retardant...

Страница 7: ...ber WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Moni...

Страница 8: ...are raised Refer to Installing the Mattress Replacement System on page 10 Entrapment May Occur WARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and...

Страница 9: ...etics Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type The oxygen tent should NOT extend below mattress support level El...

Страница 10: ...oot ends and to the frame at the center of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised NOTE For this procedure refer to FIGURE 2 1 NOTE The powered mattr...

Страница 11: ...g this equipment Invacare product manuals are available at www invacare com or your dealer NOTE Refer to the instructions provided with the side rails for the installation procedure Installing the Pow...

Страница 12: ...t 3 Perform one of the following MA90Z Perform the following steps Detail A i Squeeze and hold the tabs on the hose connectors ii Insert the hose connectors into the control unit connector iii Push th...

Страница 13: ...tripped over 1 Examine the hospital grade power cord supplied with the control unit 2 Perform one of the following If the plug is damaged Call your supplier for a replacement hospital grade cord If t...

Страница 14: ...N LED illuminates indicating that the control unit is on If the power comes on after a power outage the control unit will go through its system initializa tion routine for a few seconds and then resum...

Страница 15: ...ngles NOTE The rotation angles are and full rotation 45 Rotation Button MA95Z 1 Press the Rotation button to select Static mode or Alternating Rotation mode NOTE When the LED is on Alternating Rotatio...

Страница 16: ...tient s weight to height ratio is above average increase the pressure setting by approximately 20 MA90Z models display levels from 0 to 9 and MA95Z models display the mmHg from 6 to 32 Automatic Wirel...

Страница 17: ...is lit and L displays in the LCD Display ii Raise the bed to 45 NOTE The control unit will beep and will return to Standby iii Perform one of the following Cancel Press the Turn Angle button to exit F...

Страница 18: ...patient wound care patient turning or patient cleaning Low Air Loss Mode In this mode the mattress goes into on demand low air loss relief mode Max Inflate Button MA95Z 1 Press the Max Inflate button...

Страница 19: ...utage NOTE During a power outage the MA90Z mattress retains the air as long as the mattress is connected to the control unit Low Pressure LED In the event that the mattress hose disconnects an alarm s...

Страница 20: ...patient is lying flat on his her back in the middle of the mattress B Place four fingers between the air cushions directly underneath the sacral region of the patient s body C Ensure that there is 3 t...

Страница 21: ...to the owner s manual provided with the wheelchair 3 Press the Max Inflate button to achieve maximum mattress pressure 4 Raise or lower the bed to match the wheelchair height Refer to the owner s man...

Страница 22: ...22 Part No 1148139 FIGURE 2 5 Preparing for CPR Procedure Connector Cover Control Unit Connectors MA95Z Hose Connector DETAIL A Control Unit RED CPR Connector MA90Z Only Mattress DETAIL B DETAIL D MA...

Страница 23: ...ganisms 1 Remove the bedding 2 If necessary inflate the mattress 3 Ensure that the control unit is off 4 Unplug the power cord from the wall outlet 5 Ensure that the underside of the mattress is clear...

Страница 24: ...ings with a quaternary disinfectant solution B Using a nylon brush gently clean all crevices as they can harbor microorganisms C Air dry all treated surfaces 10 Perform the following steps to clean th...

Страница 25: ...trol unit using the two screws FIGURE 3 1 Changing the Filter Troubleshooting PROBLEM CAUSE SOLUTION Mattress not inflating Not alternating properly Mattress hose disconnected Connect hose connectors...

Страница 26: ...ck power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Check power source has power power and turn it On Power outage Wait till th...

Страница 27: ...SECTION 3 MAINTENANCE AND TROUBLESHOOTING Part No 1148139 27 MA90 Series NOTES...

Страница 28: ...ion will issue a return authorization The defective unit or parts must be returned for warranty inspection using the serial number when applicable as identification within thirty days of return author...