Invacare LiNX REM110 Скачать руководство пользователя страница 22

21

60126083-B

Powered Seat Tilt and Elevating Seat

A

 (Forward)

Powered seat tilt up

B

 (Reverse)

Powered seat tilt down

C

 (Left)

Elevating seat up

D

 (Right)

Elevating seat down

Dual-Powered Elevating Legrests

A

 (Forward)

Left powered elevating legrest up

B

 (Reverse)

Left powered elevating legrest down

C

 (Left)

Right powered elevating legrest up

D

 (Right)

Right powered elevating legrest down

4–way Button Switch

 1.  Make sure the mobility device is on a level surface and is 

turned on.

 2.  Press and hold the appropriate button to move a 

particular seating function. 

The seating function moves as long as the button is 
pressed. 

See the following tables for buttons and powered seating 
function combinations.

The following tables show the factory settings. For 
reprogramming, contact your provider.

Powered Seat Tilt and Powered Recline

A

Powered seat tilt up

B

Powered seat tilt down

C

Powered recline up

D

Powered recline down

Powered       Seat Tilt, Recline, Elevate and Legrest

A

Powered seat tilt toggle up/down

B

Powered seat recline and legs up/down 

C

Powered seat elevate up/down

Powered       Seat Tilt, Recline, Elevate and Legrest

D

Powered        seat legs up/down

Powered Seat Tilt and LNX Legrest

A

Powered seat tilt up

B

Powered seat tilt down

C

LNX up

D

LNX down

Powered Recline and LNX Legrest

A

Powered recline and LNX up

B

Powered recline and LNX down

C

LNX up

D

LNX down

Powered Seat Tilt and Elevating Seat

A

Powered seat tilt up

B

Powered seat tilt down

C

Elevating seat up

D

Elevating seat down

Dual-Powered Elevating Legrests

A

Left powered elevating legrest up

B

Left powered elevating legrest down

C

Right powered elevating legrest up

D

Right powered elevating legrest down

6.12.2  Through the Joystick

Activating the Seating Function

 1.  Press the Seating function key 

A

.

The wheelchair changes to the seating function, and the 
Drive/actuator status display 

C

 lights up amber.

LiNX® Control System

Содержание LiNX REM110

Страница 1: ...1994 UNLESS OTHERWISE SPECIFIED ALL DIMENSIONS ARE IN INCHES UNLESS OTHERWISE SPECIFIED CRITICAL QUALITY ATTRIBUTE KEY CHARACTERISTIC IDENTIFICATION PART DESCRIPTION MANUAL USER LINX REHAB LITE REMOTE...

Страница 2: ...REM210 REM211 REM215 REM216 Supplement to power wheelchair user manual en Remote User Manual This manual MUST be given to the user of the product BEFORE using this product this manual MUST be read and...

Страница 3: ...ing Displayed Symbols and Meanings 22 6 12 3 Speed Reduction and Seating Function Inhibits 22 Speed Reduction 22 6 13 Charging the Batteries 22 6 13 1 Battery Alarms 23 High Voltage Warning 23 Low Vol...

Страница 4: ...11 2 Limited Warranty Canada 35 11 3 Repair or Replacement 35 11 4 Disclaimers 35 11 4 1 Limitations and Exclusions 35...

Страница 5: ...certain work 1 3 Prescription Statement Per 21 CFR 801 109 b 1 the device is labeled for prescription use only CAUTION Federal Law USA restricts this device to sale by or on the order of a licensed ph...

Страница 6: ...e all hardware is securely tightened after setup service or adjustments Warranty is void if non qualified individuals perform any work on this product DANGER Risk of Death Serious Injury or Damage Con...

Страница 7: ...or serious injury from these fires and resulting fumes because they may not have the ability to move away from the wheelchair DO NOT smoke while using this wheelchair WARNING Risk of Injury Damage or...

Страница 8: ...unlit rooms during the night or similar situations ALWAYS be aware of motor vehicles and your surroundings WARNING Risk of Injury Damage or Death Use of the wheelchair while judgment or ability is im...

Страница 9: ...Death Serious Injury or Damage An improperly connected joystick could cause loss of power resulting in death serious injury or damage Ensure the joystick is securely connected to controller CAUTION Ri...

Страница 10: ...ing appropriate EMC standards l USA ANSI RESNA WC 2 2009 Sec 21 l Europe EN12184 2014 ISO7176 21 2009 National and international directives require confirmation of compliance on particular vehicles Si...

Страница 11: ...REM210 l Drive function l Seating function up to two actuators Non Expandable Remotes REM215 l Drive function l Seating function up to two actuators l Lights Hazard lights REM211 l Drive function l S...

Страница 12: ...E Joystick 4 1 2 User Interface REM210 and REM211 The REM210 and REM211 remotes have the same user interface The REM211 remote has a ring around the base of the joystick A ON OFF button Status indicat...

Страница 13: ...ase of the joystick A ON OFF button Status indicator B Battery gauge C Speed dial D Horn E Joystick F Drive actuator status G Connectivity indicator H Seating function selector I Lights and direction...

Страница 14: ...e l Please be environmentally responsible and recycle this product through your recycling facility at its end of life D Tamper evident seal E Product label containing l Dynamic Controls dynamic logo l...

Страница 15: ...e using the letters A to L A Jan B Feb C Mar etc l YY is the year of manufacture l nnnnnn is a unique six digit sequential number For example the remote serial number begins with A14 This indicates th...

Страница 16: ...ower This results in an unintended stop Always check the plug of the remote for damage Contact your provider immediately in case of a damaged plug Risk of Damage to the Remote The remote plug and conn...

Страница 17: ...o lock the system see 1 1 Locking unlocking the remote page 1 6 1 3 Joystick Joystick shaping is used to calibrate a joystick to reduce the extent to which users must deflect their joystick to reach f...

Страница 18: ...n or profile they are using and are especially beneficial for l users with impaired vision or users who cannot see the display l users who want additional feedback from their actions without the need...

Страница 19: ...ove the joystick if this parameter has been enabled by the provider 6 11 Using Rest Mode The rest state of the LiNX system provides the occupant with a safe environment in which to sleep rest or carry...

Страница 20: ...owing occurs l Normal operation is resumed manually l The system times out into sleep if configured to do so l The system is power cycled How normal operation is resumed manually depends on if the dri...

Страница 21: ...d is turned on 2 Press and hold the stereo buttons A or B to move a particular seating function The seating function moves as long as the button is pressed 4 way Toggle Switch 1 Make sure the mobility...

Страница 22: ...Powered seat tilt down C Powered recline up D Powered recline down Powered Seat Tilt Recline Elevate and Legrest A Powered seat tilt toggle up down B Powered seat recline and legs up down C Powered s...

Страница 23: ...reduction is shown in the status display If the elevating seat is raised above a certain point the elevating seat symbol and the drive symbol pulse These two symbols continue pulsing while driving to...

Страница 24: ...For more information about the charging state refer to the charger user manual or the power wheelchair base user manual While charging the wheelchair is in drive inhibit mode The drive inhibit mode en...

Страница 25: ...tic or unintended movement of the wheelchair may occur if wireless transmitters are connected to the wheelchair To avoid injury or damage DO NOT use the USB charger connector as a wireless transmitter...

Страница 26: ...icing Weekly monthly and periodic inspections should be performed by the user attendant between six month service inspections Regular cleaning will reveal loose or worn parts and enhance the smooth op...

Страница 27: ...age Cleaning or maintenance may cause damage to carpeting or flooring Place the wheelchair in a well ventilated area where cleaning or maintenance can be performed without risk of damage to carpeting...

Страница 28: ...indication and a few possible actions that can be taken to rectify the problem The actions listed are not in any particular order and are suggestions only The intention is that one of the suggestions...

Страница 29: ...powering up l On a function change l When transitioning from a drive lockout or inhibit state Otherwise a drive OON warning displays During a drive OON warning the battery gauge LEDs and the drive whe...

Страница 30: ...until the fault condition is cleared 8 4 Cut off Voltage When the battery voltage decreases below the battery cut off voltage l The status indicator flashes red flash code 2 refer to 1 1 Fault codes...

Страница 31: ...for storage according to ISO 7176 9 40 F 40 C 149 F 65 C Operation humidity range according to ISO 7176 9 0 90 RH Degree of protection IPX41 1 IPX4 classification indicates the electrical system is p...

Страница 32: ...ser defined Robustness Adaptive fast frequency hopping FEC fast ACK Adaptive frequency hopping Lazy Acknowledgment 24 bit CRC 32 bit Message Integrity Check Latency from a non connected state Typicall...

Страница 33: ...te control of the wheelchair remains with the end user at all times Should a user determine during the customization process that the wheelchair setup is inappropriate in providing full control in eve...

Страница 34: ...device during use remove the wheelchair from the RF field and wirelessly reconnect the device 10 6 Cybersecurity The LiNX product range has been designed with cybersecurity in mind to assure device fu...

Страница 35: ...rsecurity of the LiNX system However should the user be concerned about the Bluetooth connection for any reason the user can switch off the Bluetooth functionality by powering down the system The user...

Страница 36: ...coverage under the Warranty 11 3 Repair or Replacement Invacare s sole obligation and the original purchaser s exclusive remedy under the Warranty is limited to Invacare s repair and or replacement at...

Страница 37: ...failure to adhere to the product instructions A change in operating noise particularly relative to motors and gearboxes does not constitute a failure or defect and will not be repaired or replaced as...

Страница 38: ...37 60126083 B Notes LiNX Control System...

Страница 39: ...60126083 B 38 Notes 11 Warranty...

Страница 40: ...39 60126083 B Notes LiNX Control System...

Страница 41: ...m Canada 570 Matheson Blvd E Unit 8 Mississauga Ontario L4Z 4G4 Canada 800 668 5324 www invacare ca Manufacturer Invacare Corporation One Invacare Way Elyria OH 44035 USA Dynamic Controls 39 Princess...

Отзывы: