SECTION 4—INSPECTION AND COMPRESSOR INTERFACE

Part No 1112648

13

HomeFill

FIGURE 4.1   

External Examination

Improper filling or abuse of this cylinder, or failure to follow these safety
instructions may cause serious bodily injury or death. High pressure
oxidizing gas vigorously accelerates combustion.

DO NOT alter or modify cylinder or related components.
DO NOT use oil or grease. Keep cylinder, cylinder valve, all equipment 
and connections free of oil and grease to avoid violent ignition resulting 
in serious bodily injury or death.

For medical applications. Use only as directed by a licensed 
practitioner. Uninterrupted use of high concentrations of oxygen over 
a long duration, without monitoring its effect on oxygen content of 
arterial blood, may be harmful.

Oxygen Cylinders filled by the Invacare HomeFill oxygen 
system shall be used for personal use only. 

"

Not to be 

filled for resale or use by professional users.

"

This cylinder is to be used with the Invacare HomeFill 
oxygen system. This cylinder contains 93 ± 3% oxygen 
produced by an oxygen concentrator using the pressure swing 
adsorption process.

DO NOT handle cylinder or use contents until you are professionally 
trained, including emergency procedures.

Use in accordance with Invacare HomeFill oxygen system Operator’s 
Manual. Open valve slowly. Close valve after each use and when 
empty. Secure cylinder during storage and use. No smoking in cylinder 
area. Keep away from heat, flame and spark. Keep out of reach of 
children. DO NOT drop.
Filling of this gas is performed by Invacare HomeFill oxygen system 
Only.
Cylinders that have been refinished and/or exposed to elevated
temperatures must be hydrostatically tested before filling. Cylinder 
temperature should not exceed 125°F (52°C). Aluminum cylinders 
subject to the action of fire or heated to temperatures in excess of 
350°F must be withdrawn from service and condemned by trained 
personnel. DO NOT use caustic paint strippers. Valve and safety 
relief devices must be removed and replaced by trained and 
authorized personnel. DO NOT alter or change valve assembly.

DO NOT REMOVE THIS PRODUCT LABEL

Not for Commercial Resale

Part Number 1125090   Rev D

Invacare Corporation
2101 E. Lake Mary Blvd.
Sanford, FL 32773
1-800-832-4707

Содержание HomeFill HF2PC9

Страница 1: ...ser Before using this cylinder read this manual and save for future reference For more information regarding Invacare products parts and services please visit www invacare com Operator s Manual HomeFill Oxygen Conserving Cylinder HF2PCL6 HF2PC9 HF2PCL4 ...

Страница 2: ... ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS OR INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR TECHNICAL PERSONNEL BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY OCCUR ACCESSORIES WARNING Invacare products are specifically designed and manufactured for use in conjunction with Invacare accessories Accessories designed by other manufacturers have not been tested ...

Страница 3: ...enance 8 SECTION 2 FEATURES 9 SECTION 3 TYPICAL PRODUCT PARAMETERS 10 SECTION 4 INSPECTION AND COMPRESSOR INTERFACE 11 Cylinder Prefill Inspection 11 Connecting Disconnecting the Cylinder to from the Compressor 14 SECTION 5 OPERATING INSTRUCTIONS 19 SECTION 6 MAINTENANCE 21 SECTION 7 CYLINDER FILL TIMES 22 Troubleshooting 22 Replacement Parts Accessories 22 LIMITED WARRANTY 23 ...

Страница 4: ...here oxygen is being delivered NO SMOKING signs should be prominently displayed Textiles and other materials that normally would not burn are easily ignited and burn with great intensity in oxygen enriched air Failure to observe this warning can result in severe fire property damage and cause physical injury or death SIGNAL WORD MEANING DANGER Danger indicates an imminently hazardous situation whi...

Страница 5: ...om package and inspect for damage If there is any damage DO NOT use Contact your dealer Home Care equipment provider INTENDED USE To store regulate and deliver supplemental oxygen to patients who may have difficulty extracting oxygen from the air they breathe It is for patients who would normally receive the oxygen via a nasal cannula The device delivers USP 93 oxygen at a number of different flow...

Страница 6: ...oduct contains magnetic ferrous material that may affect the results of an MRI Be sure all connections are tight and leak free DO NOT autoclave DO NOT gas sterilize with ethylene oxide DO NOT clean with aromatic hydrocarbons DO NOT immerse product in any kind of liquid Store product in a clean area when not is use Only use Invacare Corporation HF2PCL6 carrying bag HF2PC9 carrying bag or HF2PCL4 ca...

Страница 7: ...hands properly prior to usage DO NOT allow oil from your hands or other sources to come into contact with the compressor coupling or bottle fittings These substances may become flammable in the presence of oxygen and cause injury Avoid touching the bottle fitting whenever possible Example of items to avoid are hand creams or lotions cooking oils suntan oils or sunblocks and similar oily products I...

Страница 8: ... DO NOT use while sleeping without consulting your healthcare provider This product is designed to operate with a standard single lumen adult oxygen nasal cannula with a maximum length of seven feet The cannula is for single patient use only Cannula MUST be installed on outlet connection prior to use DO NOT place the unit under clothing while in use When this product is in use a small amount of ox...

Страница 9: ...rt No 1112648 9 HomeFill SECTION 2 FEATURES DANGER This product is not intended as a life sustaining or life supporting device FIGURE 2 1 Product Diagram Outlet Connection Cannula Flow Dial Cylinder Gauge Cylinder Fill Port ...

Страница 10: ...Overall Length 14 2 in 36 cm 14 in 35 5 cm HF2PC9 Diameter 4 4 in 11 1 cm 4 4 in 11 1 cm HF2PCL4 Weight N A 3 6 lbs 1 63 kg HF2PCL4 Overall Length N A 8 9 in 22 2 cm HF2PCL4 Diameter N A 4 4 in 11 1 cm PULSE SETTINGS 1 2 3 4 5 FLOW SETTINGS 2 LPM Continuous SAVINGS RATIO Up to 4 3 1 Up to 5 7 1 CANNULA REQUIREMENT 4 ft recommended MS0204 7 ft long max std adult single lumen oxygen nasal cannula OP...

Страница 11: ...tempting to fill otherwise injury or damage may occur NEVER smoke in an area where oxygen is being administered NEVER use near any type of flame or flammable explosive substances vapors or atmosphere NOTE For this procedure refer to FIGURE 4 1 The Hydrostatic Testing Date Aluminum cylinders MUST undergo testing every five years WARNING DO NOT fill cylinders that have not been tested in the past fi...

Страница 12: ...t cause a cylinder to be unacceptable or unsafe for use 2 Examine the cylinder for evidence of fire or thermal damage Evidence includes charring or blistering of the paint or other protective coating or heat sensitive indicator If fire or thermal damage is found replace the cylinder 3 Inspect the cylinder fillport for the following and replace if found Debris oil or grease Noticeable signs of dama...

Страница 13: ... 3 oxygen produced by an oxygen concentrator using the pressure swing adsorption process DO NOT handle cylinder or use contents until you are professionally trained including emergency procedures Use in accordance with Invacare HomeFill oxygen system Operator s Manual Open valve slowly Close valve after each use and when empty Secure cylinder during storage and use No smoking in cylinder area Keep...

Страница 14: ... to test for leaks NOTE For this procedure refer to FIGURE 4 2 and FIGURE 4 3 1 Examine the cylinder gauge If the cylinder pressure is less than 1500 psig proceed to STEP 2 FIGURE 4 3 NOTE If the cylinder pressure is greater than 1500 psig DO NOT attempt to top off this cylinder It may not fill 2 Set the regulator flow dial on the cylinder to OFF FIGURE 4 3 3 Remove the cylinder and connector fill...

Страница 15: ...der regulator assembly in the area behind the cylinder gauge FIGURE 4 3 6 Position the cylinder in the compressor cradle FIGURE 4 3 7 Align the cylinder fillport with the connector fillport FIGURE 4 3 8 Pull UP on the outer ring sleeve of the connector fillport while pushing DOWN on the cylinder regulator assembly to couple the cylinder fillport into the connector fillport FIGURE 4 3 NOTE The cyli...

Страница 16: ...r Cylinder Regulator Assembly Compressor Cylinder Gauge Cylinder Fillport Connector Fillport PULL UP Outer Ring Sleeve Cylinder Compressor Cradle Cylinder Gauge Regulator Flow Dial Outer Ring Sleeve Connector Fillport Outlet Cannula Connection Power Switch Regulator Cylinder Fillport NOTE Fillport covers not shown for clarity ...

Страница 17: ...IGURE 4 4 1 Press the compressor power switch to the Off O position 2 Grasp the cylinder regulator assembly in the area behind the cylinder gauge 3 With the other hand grasp the outer ring sleeve of the connector fillport and push DOWN 4 Lift up on the cylinder regulator assembly to remove from the connector fillport FIGURE 4 4 Disconnecting the Cylinder from the Compressor Cylinder Regulator Asse...

Страница 18: ...connected from the connector fillport release the outer ring sleeve of the connector fillport and use two hands to remove the cylinder from the compressor cradle 6 Place the fillport covers onto the connector and cylinder fillports NOTE The fillport covers should be used whenever cylinders are not being filled ...

Страница 19: ...your dealer Home Care equipment provider 2 Be sure regulator is in the Off position Align Off with the indicating pointer 3 Position the cylinder so the oxygen cylinder valve outlet is pointing away from the user and any other person s WARNING DO NOT direct flow of oxygen at any person or flammable material when adjusting the flow dial 4 Attach a standard adult single lumen oxygen nasal cannula no...

Страница 20: ... To obtain desired flow the indicating pointer must align with a specific number on the dial To avoid injury to patient ALWAYS confirm prescribed setting before administering to patient and monitor flow on a frequent basis Only use Invacare Corporation HF2PCL6 carrying bag the HF2PC9 carrying bag or the HF2PCL4 carrying bag 7 Breathe through the nose and feel a pulse of oxygen as you inhale NOTE W...

Страница 21: ... use clean exterior of the product with a dry lint free cloth only 2 Store product in a clean area free from grease oil and other sources of contamination CAUTION DO NOT use cleaning solutions DO NOT immerse product in any kind of liquid All repairs MUST be done by Invacare Corporation ...

Страница 22: ...Carrying Bag or HF2PCL4 Carrying Bag Cylinder Filling Times Concentrator Flow Rate to Patient Platinum 5 up to 3 L min and Platinum 10 up to 5 L min HF2PC9 HF2PCL4 HF2PCL6 NORMAL 2 hr 20 min 60 min 1 hr 25 min MAXIMUM 2 hr 45 min 1 hr 20 min 1 hr 50 min PROBLEM POSSIBLE CAUSE SOLUTION No flow 1 Regulator in Off position 1 Set to prescribed setting 2 Cylinder empty 2 Replace fill cylinder 3 Dial no...

Страница 23: ... this manual Upon receiving notice of an alleged defect in a product Invacare will issue a serialized return authorization It shall be the responsibility of the purchaser to pack the product s or part s in an appropriate manner to avoid shipping damage and return the product s or part s at the purchaser s expense to either Invacare s plant or service center as specified by Invacare in advance The ...

Страница 24: ...DENTAL OR CONSEQUENTIAL DAMAGE OR LIMITATION OF HOW LONG AN IMPLIED WARRANTY LASTS SO THE ABOVE EXCLUSION AND LIMITATION MAY NOT BE APPLICABLE THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Invacare Corporation www invacare com USA One Invacare Way Elyria Ohio USA 44036 2125 440 329 6000 800 333 6900 Technical Services 440 329 6593 800 832 4707 2011 Invacare ...

Отзывы: