SECTION 3—INSTALLING THE CONSERVING REGULATOR
13
NOTE:
For
information
on
how
to
operate
the
cylinder,
consult
your
equipment
provider.
NOTE:
For
this
procedure,
refer
to
1. Loosen
the
cylinder
post
valve
T
‐
Handle.
2. Lower
the
conserver
over
any
CGA870
post
valve
cylinder
with
the
alignment
pins
toward
the
holes
on
the
cylinder
neck.
3. Align
the
two
pins
and
gasket
with
the
corresponding
holes
on
the
cylinder
post
valve.
FIGURE 3.1
Attaching the
IOC100P to the Cylinder
NOTE:
Cylinder
post
valve
T
‐
Handle
should
be
aligned
with
indentation
on
post
valve.
4. While
holding
the
unit
in
place,
tighten
the
cylinder
post
valve
by
turning
clockwise.
5. Attach
a
standard
cannula
[7
ft.
(2.13
m)
or
less
in
length]
to
the
oxygen
supply
outlet
fitting.
NOTE:
Tighten
only
by
hand.
The
use
of
a
tool
to
tighten
the
T
‐
Handle
may
damage
the
unit.