46
7.8.1
Error codes and diagnostic codes
The drive electronics are capable of rectifying some errors automatically. In this case the status
display will cease to flash. Please switch the remote on and off several times. Wait approx. 5
seconds each time before switching the remote on again. If this does not rectify the error, locate
the error using the flash codes shown below.
FLASH
CODE
FAULT
IMMEDIATE MEASURE
FURTHER HELP
1
Module defective.
-
•
Consult dealer.
2
Lifter raised or
lowered too far (seat
not at driving height)
•
If lifter is raised, lower in
stages until the status
display stops flashing. If
lowered too far, raise lifter
in stages until the status
display stops flashing. If at
all possible, only drive
when the seat is at driving
height.
-
Accessory error.
-
•
Consult dealer.
3
Error on right motor.
Connection
loose/faulty.
•
Check plug-in connectors.
•
Consult dealer.
4
Error on left motor.
Connection
loose/faulty or motor
faulty.
•
Check plug-in connectors.
•
Consult dealer.
Содержание Electric wheelchair
Страница 1: ...Invacare Typhoon Electric wheelchair Operating manual ...
Страница 55: ...55 Loosen the counternut 1 on the spindle Set the seat angle by turning the spindle 2 Tighten the counternut ...
Страница 58: ...58 8 4 2 Flex and Contour Seats Set the angle by turning the hand wheel 1 ...
Страница 74: ...74 Use the 6mm Allen key to loosen and remove the screws 1 on both sides Pull cover to the front and remove ...
Страница 77: ...77 Release and remove the battery cable on the battery poles Remove batteries by pulling to the front ...
Страница 97: ...97 Order No of this Manual 1435058 DOC Release Date 01 03 04 English ...