45
7.8 Error
diagnosis
In the event that the electronics should show signs of failure, please consult the following
troubleshooting guide in order to locate the error.
INFORMATION
Before beginning with the diagnosis, please ensure that the drive electronics are switched on.
If the status display is OFF:
Please check whether the drive electronics are SWITCHED ON.
Please check whether all cables have been connected correctly.
Please ensure that the batteries are not discharged.
If the status display is FLASHING:
Please count the number of flashing sequences and move on to the next section.
If the red diodes on the battery charger display and the status display are FLASHING,
Drive mode display shows a horizontal bar:
Battery discharged. Please charge the battery.
Содержание Electric wheelchair
Страница 1: ...Invacare Typhoon Electric wheelchair Operating manual ...
Страница 55: ...55 Loosen the counternut 1 on the spindle Set the seat angle by turning the spindle 2 Tighten the counternut ...
Страница 58: ...58 8 4 2 Flex and Contour Seats Set the angle by turning the hand wheel 1 ...
Страница 74: ...74 Use the 6mm Allen key to loosen and remove the screws 1 on both sides Pull cover to the front and remove ...
Страница 77: ...77 Release and remove the battery cable on the battery poles Remove batteries by pulling to the front ...