The motor can be turned to four different positions:
1.
Remove bolt/pin
A
from end of motor
B
unit.
2.
Rotate motor
B
unit to the desired position and carefully
replace the bolt/pin
A
.
Mounting Motor on the Opposite Side
The motor can be mounted either on the left or on the right
side of the Backrest Support.
1.
Remove the Backrest Support from the bed.
2.
Using the hand control, raise Backrest Support to the
highest position.
3.
Remove the pivot joint protectors
A.
4.
Place the Backrest Support on side opposite of motor.
5.
Loosen screw
B
with 6 mm Allen wrench
C
. Loosen
screw only as much as necessary to rotate arm freely.
6.
Rotate arm 180 degrees.
7.
Reverse steps 3-5 to reassemble Backrest Support.
8.
Reinstall Backrest Support onto bed. Refer to
Backrest Support on Bed, page 2
Accessory Bag
The optional accessory bag is offered for storing the hand
control and power supply for the Invacare Bedco Backrest
Support.
Maintenance
Cleaning
WARNING!
Risk of Injury or Death
To avoid electrocution:
– Do not perform service or maintenance on any
electrical parts on this product unless properly
trained.
– Do not use water or lubrication on any electrical
components.
IMPORTANT!
To avoid damaging the product:
– DO NOT use abrasive cleaners to clean the
canvas cover.
•
Canvas cover is machine washable at a max temperature
of 60°C. See attached label on the textile cover for
detailed washing instructions.
•
With the exception of electronic components, other
parts on the Backrest Support can be washed using a
damp sponge or cloth.
Lubrication
If necessary, pivot joints can be sprayed with a silicone based
lubricant.
After Use
Waste disposal
Please be environmentally responsible and recycle this
product through your recycling facility at its end of life.
Waste disposal must comply with the laws and regulations
pertaining waste handling in each country.
Reuse
WARNING!
Risk of Injury
To avoid injury when transferring the product to a
different user:
– Always have a dealer inspect the product for
damage prior to transfer.
– If damage is found, do not use the product.
Instead contact dealer or Invacare for service.
Technical Data
Product Parameters
Mechanical Parameters
Bedco
750
Bedco
900
Bedco
1050
Bedco
1200
Total width [mm]
800
950
1000
1200
Backrest width [mm]
700
850
1000
1200
Backrest Angle
70°
Weight [kg]
9.5
10
10.5
13
Maximum load
1
[kg]
70
77
Maximum user weight
[kg]
135
150
1
On the middle of the backrest support.
Electrical Parameters
Voltage Supply
230 V AC ±10%, 50 Hz
Max Current Input
1 A
Intermittent
(Periodic Motor Operation)
10%, Max. 2 min. / 18 min.
Protection Class
1PX4
Insulation Class
Class II
Protection
Type B Applied Part
1417684-I