Invacare® Action 3 NG
•
Swinging calf pad forwards (1) or backwards (2) :
1.
Swing calf pad
E
forwards (1).
2.
Pull up the calf pad
E
and swing it backwards (2).
•
Adjusting height of calf pad:
1.
Loosen knob
C
.
2.
Adjust to the desired height and firmly tighten the
knob.
•
Adjusting depth of calf pad
The calf pad has four depth adjustment options:
1.
Remove the fixing screw
F
with 5 mm Allen key.
2.
Adjust to one of the four positions and firmly
tighten the fixing screw
F
(3–5 Nm).
Adjusting foot plate
There are three different fold-up foot plates available.
•
Standard height-adjustable foot plates:
1.
Loose the fixing screw
B
with 5 mm Allen key.
2.
Adjust the height and let the screw catch one of
the recesses on the foot plate tube.
3.
Tighten the fixing screw
B
(3–5 Nm) in the desired
position.
The distance between the lowest part of the
footrest and the ground must be at least 50 mm.
•
Depth-, Angle- (1) and Width- (2) adjustable foot plates:
1.
Loosen the fixing screw
G
or
H
with 5 mm Allen
key.
2.
Adjust the depth and angle positions and firmly
tighten the fixing screw
G
(12 Nm) or
H
(8–9 Nm)
in the desired position.
28
1608467-E