SYMPTOM:
PROBABLE CAUSE:
SOLUTION:
NOTE: REFER TO THE REPAIR/REPLACEMENT GUIDE IN THIS MANUAL FOR PROPER
REPLACEMENT OR ADJUSTMENT PROCEDURES
NORMAL OPERATION:
Internal Status Indicators:
RED: Off GREEN: Off
Unit plugged in, power switch on.
3 "beeps" on start up.
POWER LOSS:
Internal Status Indicators:
RED: Off GREEN: Off
Unit unplugged, power switch on,
alarm off.
No problems.
POWER LOSS:
Internal Status Indicators:
RED: Off GREEN: Off
Unit plugged in, power switch on,
alarm off. "Battery Free
TM"
circuit
drained.
1. No Power at outlet.
2. Power Cord.
a. Frayed.
b. Broken or damaged spade.
c. Spade connector from Power
Cord loose or disconnected
(inside back of unit).
3. On/Off switch.
a. Disconnected wire.
b. Faulty switch.
1. Check electrical outlet with a
table lamp or voltmeter set on
150-200 VAC scale. If outlet isn't
working, check protective device
in home's electrical panel or
consult an electrician. Also
ensure that unit is properly
plugged in. DO NOT use
extension cords. Move to another
outlet or circuit.
2a.Replace cord.
2b.Replace power cord connectors
on plug.
2c.Reattach connector.
3a.Check all electrical connections
to the On/Off switch for any
disconnected wires.
3b.If the concentrator does not come
on at all and wiring is intact, color
code and remove wires one at a
time and clean the On/Off switch
with contact cleaner while turning
the switch on and off. Reinstall
switch and turn power on. If
switch still doesn't work, replace
switch. Transfer wires from old
switch to new switch one at a time
to the matching contact.
No problems.
System O.K.
TROUBLESHOOTING GUIDE
TROUBLESHOOTING GUIDE
"Battery Free
TM
" Circuit drained. Plug
in cord and turn power switch ON to
recharge.