Intersurgical PMH7000 Скачать руководство пользователя страница 18

Instruction For Use                                                           PMH7000                                                         

       

Version 2

   

 
 

Revision 03                                                                  23 July 2019    

 

                                                   Page 18 of 20 

10.  Reprocessing 

PMH7000 humidifier bases and accessories should be cleaned, disinfected and sterilized according to the instructions in 
the Cleaning, Disinfection and Sterilization Manual before use. 
 

Components and 
Parts 

Material 

Autoclave 
Sterilization 

Cleaning and Disinfectant  
Solution 

ETO Gas 
@ 55

°

Dual Temperature 
Probe 

Housing: Polycarbonate 
 

No 

Yes 

Yes 

Base Unit 

Housing: Polycarbonate 

No 

Yes 

No 

ETO Gas Sterilization: Sterilization Time: 4 hours at 55

°

C,  Aeration Time: min. 15 hours 

 

Note: 

CaviWipe

TM

 (or equivalent) Disinfecting Towelettes to clean and disinfect all surfaces of the Humidifier base and Dual 

Temperature probe. 
 

Note: 

Ensure that the heater plate surface is clean and free from excessive abrasions. 

 
Warning !  

DO NOT IMMERSE or STERILIZE main unit of PMH7000 Humidifier heater base. 

Warning !  

DO NOT STERILIZE or AUTOCLAVE the heater base unit. 

 

11.  Repair 

Repair of the unit should be done only by a qualified technician. 
Please contact your sales provider if there are any problems. 

 

12.  Disposal 

Some authorities consider these materials are subject to special disposal.  
Contact your local authority for the detailed information.  
 

13.  Trouble Shooting 

The device has an alarm function, which shows any abnormal conditions during operation. 
Each alarm shows the abnormal conditions by illuminating an LED on the front panel and sounding the alarm  
Alarm sound can be muted by the mute button for 2 minutes, but the LED will keep flashing until the alarm condition itself is 
removed. 

LED Indicator

 

Alarm 

Sound 

Possible Cause 

User Action 

Patient Icon 
(Abnormal high 
temperature: Flashes

 

red) 
 

Yes 

1) 

Patient temperature

set value  

2°C 

 

2) No increase over 

2°C in two 

minutes after power on 

Confirm first that there is no problem 
on setting and breathing circuit.  If the 
alarm conditions are not restored, stop 
using the unit and contact your sales 
provider 

Patient Icon 
(Abnormal low 
temperature :Flashes

 

yellow) 

Yes 

1) Patient temperature

set value 

 -4

 

Check dual temperature probe, and 
restore the alarm condition. 

Chamber Outlet Icon 
(Abnormal high 
Temperature: Flashes

 

red) 

Yes 

1)

  Chamber outlet 

temperature

set value+ 4°C 

lasting for more than 20 
minutes. 

2)

  Chamber outlet temperature 

reaches 10°C higher than set 
temperature.  

 

Confirm that there is no problem with 
the setting.   Review alarm conditions 
and restore. If the alarm conditions are 
not restored, stop using the unit and 
contact your sales provider for repair. 

Chamber Outlet Icon  
(Abnormal low 
Temperature: Flashes

 

yellow) 

Yes 

1)  Chamber temperature

set 

value -4°C lasting for more than 
20 minutes.  

2)  Chamber temperature reaches 

10°C lower than the set 
temperature. 

Confirm the conditions of dual 
temperature probe, and restore the 
alarm condition. 

Содержание PMH7000

Страница 1: ...03 23 July 2019 Page 1 of 20 Intersurgical Humidifier Heater Base PMH7000 Instructions for Use Distributed by Intersurgical Ltd Crane House Molly Millars Lane Wokingham Berkshire RG41 2RZ UNITED KING...

Страница 2: ...late Temperature 7 3 1 8 Heater Wire Resistance 7 3 2 Physical Dimensions 7 3 2 1 Size Without Humidification Chamber Installed 7 3 2 2 Weight 7 3 3 Environmental 7 3 3 1 Operating Temperature 7 3 3 2...

Страница 3: ...Left Side View 14 6 2 Left Side Panel Description 14 7 ASSEMBLY WITH BREATHING CIRCUIT 15 8 SET UP AND OPERATION OF THE HUMIDIFIER BASE 16 9 MAINTENANCE REPAIR 17 9 1 Basic Checks 17 9 2 Dual Tempera...

Страница 4: ...vice 2 1 Warnings Cautions and Notes In this manual a WARNING describes the actions necessary to prevent injury or loss of life A CAUTION describes the actions to be taken to protect the equipment fro...

Страница 5: ...n cause patient injury if not acted upon WARNING ALARMS No alarms sound or alarm location lights on the front panel can be activated while the system is shut down or while the protective thermostat is...

Страница 6: ...ing Radiation Power On Off IPPV Mode NPPV Mode Alarm Silencer Button Do not discard 3 Specifications WARNING EXPLOSIVE HAZARD Do not use this equipment in the presence of flammable gases or anaestheti...

Страница 7: ...10 to 95 Non condensing That includes the transportation time 3 3 5Atmospheric Pressure 700 to 1060 hPa 3 4 Electromagnetic Interference PMH7000 complies with limitations as specified in IEC 60601 1 2...

Страница 8: ...y network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Class A Voltage fluctuations flicker emissions IEC 61000 3 3 Complies Guidance and manufacturer s declarat...

Страница 9: ...ordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess electromagnetic environment due to fixed RF...

Страница 10: ...0 Humidifier base incorporates a shunt fuse the PMH7000 is not compatible for use with electrical adaptors containing a shunt fuse 3 6 Settings for Ventilated Patients IPPV mode When the gas flow is i...

Страница 11: ...est and alarm sound test as well When it is turned off alarm sound verification will be activated only once in short time Note In order to verify the functionality of the alarm system during the power...

Страница 12: ...nk quickly in green and display patient temperature setting At this time temperature setting changes can be made by pressing up and down button Press the button once and the temperature value changes...

Страница 13: ...rmittently LED Dual temperature probe Flashes red The dual temperature probe 7700000 is disconnected or faulty in this case a high priority audible alarm sounds intermittently LED Heater plate Flashes...

Страница 14: ...Connector Temperature probe connecting port 3 Inspiratory Expiratory Heater Wire Connector Heated wire adaptor connecting port NOTE The power cord cannot be detached from the heater base The power co...

Страница 15: ...g port Patient side temperature sensor Connect to the patient side temperature sensor connecting port on the breathing circuit Heater wire connection Inspiratory side Connect the connector of inspirat...

Страница 16: ...connect the cloverleaf connector to inspiratory heated wire limb and oval connector to the expiratory heated limb When using a single heated wire systems connect the electrical adaptor 5600000 to the...

Страница 17: ...n the dual temperature probe is exchanged or replaced be sure to turn off the power of the Humidifier base After exchange or replacement re activate the Humidifier base 12 Safe Process to stop operati...

Страница 18: ...abnormal conditions during operation Each alarm shows the abnormal conditions by illuminating an LED on the front panel and sounding the alarm Alarm sound can be muted by the mute button for 2 minute...

Страница 19: ...e inter lock is activated It is locked when alarm state continues for 10 minutes Turn off the device and restart it The device is not powered on No 1 Device fused 2 Heater base fault 3 Overheat protec...

Страница 20: ...Instruction For Use PMH7000 Version 2 Revision 03 23 July 2019 Page 20 of 20...

Отзывы: