Instruction For Use PMH7000
Version 2
Revision 03 23 July 2019
Page 17 of 20
(7)
Turn on PMH7000 power button.
Heater base displays “88.8” and all LED’s are activated, alarm sounds for 2 seconds then will revert to
display the previous setting for temperature (patient side and chamber side). When it does not function
as described, turn off the power and remove heater base from breathing circuit and contact your Sales
provider for assistance.
A short alarm sound will be activated when Humidifier base is turned off.
Warning !
◇
Do not use a heater base that does not properly function until probable causes are resolved.
◇
Do not use device without gas flow. If gas flow is interrupted or significantly decreased, turn
off the device until the gas flow has resumed and stabilised.
(8)
Select desired mode
When “MANUAL” mode is selected, the factory setting is 37
C at both patient side and chamber side.
This value can be changed by following instructions, please refer to Section 5.3.3 for how to change
the setting.
(9)
Connect to patient once the temperature has stabilized.
Humidification will start approx. 10 minutes later with the entire warm up period taking 30 minutes to
reach the setting temperature.
The portion physically touching to the patient (applied part) is a breathing circuit.
(10)
Remove chamber
Chamber can be easily removed by pressing the lock bar as
illustrated.
(11)
Exchange or replace the dual temperature probe
When the dual temperature probe is exchanged or replaced, be sure to turn off the power of the
Humidifier base. After exchange or replacement, re-activate the Humidifier base.
(12)
Safe Process to stop operation
Please follow below process for the operator to stop operation safety.
1) Remove the breathing circuit from the patient.
2) Press the power switch of the heater base for more than one (1) second to turn off the power.
3) Remove the power plug of the heater base from AC power supply.
4) Turn off the power switch of the gas supplying unit
.
9. Maintenance & Repair
9.1 Basic Checks
Visual inspection of the heater base and accessories should be performed monthly. Check the temperature probe sensors,
temperature sensor cables, all unit cables and connectors, and the power cord. Damaged items should be replaced if
necessary. For complete maintenance information, contact your sales provider.
9.2 Dual Temperature Probe Check
It is recommended that the dual temperature probe be functionally tested on a routine basis (every 6 months) to ensure
that it is operating properly.
The following method is recommended:
1
Obtain a precision thermometer or other temperature measuring device accurate to /-0.5
°
C.
2
Place the thermometer (or equivalent device) and both of the temperature sensors into a container of
warm water (between 25
°
C and 45
°
C).
3
Allow time for equilibration of the thermometer and the dual temperature probe. Connect the dual
temperature probe to the heater base as recommended and turn the unit on. Compare temperature
readings. The display should be /-1.5
°
C of the thermometer. If not, probe should be replaced.