Intersil ISL76683 Скачать руководство пользователя страница 4

Application Note 1657

4

AN1657.0

August 12, 2011

Using the ISL76683 Light Sensor 

Evaluation Software

USB Communications

In the upper left corner of the ISL76683 Light Sensor Evaluation 
Software window, in the USB Comm area, ensure that the light is 
green (Figure 5). If it is not green, check the USB cable 
connection.

Test Communications

In the upper left corner of the ISL76683 Light Sensor Evaluation 
Software window, in the Device Mode Control dialog area, click 
the Test Comm radio button to test communications between the 
PC and the evaluation board. If it shows Good, as shown in 
Figure 6, then the hardware and software are properly set up. If it 
shows Fail, check the connections between the PC and the 
board. If the problem persists, restart the software.

Device Registers

The Device Registers area (Figure 7) in the bottom center of the 
ISL76683 Light Sensor Evaluation Software window displays the 
current state of the device registers (Figure 7A). Use the column 
of buttons to the left of the Collect/Graph Real Time Data grid 
(top center) (Figure 7B) to read or write to the registers.

Command Register 00 (Hex)

ADC RESET AND DEVICE POWER-DOWN

In the upper left corner of the ISL76683 Light Sensor Evaluation 
Software window, in the Command Reg - b000 dialog area, click 
the ADC & Device Power Down check box to disable and reset the 
ADC and to put the ISL76683 into power-down mode.

MEASUREMENT MODE SELECTION

The ISL76683 contains two photodiodes. Diode 1 is sensitive to 
both visible and infrared light, while Diode 2 is sensitive mostly to 
infrared light. Measurement Mode 1 is Diode 1 only, and 
Measurement Mode 2 is Diode 2 only. Measurement Mode 3 is a 
sequential Mode 1 and Mode 2, with an internal subtract 
function (Diode 1 - Diode 2). 
You can select the measurement mode in one of two ways. In the 
upper left corner of the ISL76683 Light Sensor Evaluation 
Software window, in the Device Mode Control dialog area, click 
the Measure radio button, and from the scroll list, select the 
diode to be measured (Figure 9A). Or, in the Command Reg - 
b000 dialog area, under Measurement, click one of three radio 
buttons to select the diode to measure (Figure 9B). 

AD RESOLUTION

Changing the number of clock cycles does more than just change 
the resolution of the device; it also changes the integration time, 
which is the period the device’s analog-to-digital converter (ADC) 
samples the photodiode current signal for a lux measurement. To 
change the device resolution (and integration time), in the 
Command Reg - b000 dialog area, under AD Resolution, click 
one of the four radio buttons to select the number of clock cycles 
per conversion.

FIGURE 7A.

FIGURE 7B.

FIGURE 7. DEVICE REGISTERS STATUS, READ, AND WRITE

FIGURE 5. USB COMMUNICATIONS INDICATOR

FIGURE 6. TEST COMMUNICATIONS

FIGURE 9A.

FIGURE 9B.

FIGURE 9. MEASUREMENT MODE SELECTION

FIGURE 8. ADC RESET AND DEVICE POWER-DOWN 

FIGURE 10. AD RESOLUTION SELECTION

Содержание ISL76683

Страница 1: ...s 98 NT 2000 XP USB Port Hardware Setup The PC and printed circuit board PCB should not be connected via the USB until after the evaluation software has been installed satisfactorily see Software Setup on page 2 To set up the hardware 1 Connect the USB 2 0 cable first to the PC and then to the evaluation board The USB cable is the only connector needed 2 If a dialog appears asking about installing...

Страница 2: ...ntroller board ISLUSBCEVAL1Z REV A The firmware allows the evaluation software to communicate with the ISL76683 evaluation board through a PC operating on Windows 98 NT 2000 XP Installing the Software Download and run Intersil_ALS_ISL76683_Installer_V100 exe The dialog shown in Figure 2 appears Follow the instructions on each screen for installing the software The PC and PCB should not be connecte...

Страница 3: ...76683 Intersil_ALS_ISL76683 2 Double click the program name The Intersil Light Sensor Demo window shown in Figure 3 appears 3 In the Device Select dropdown menu select ISL76683 4 The ISL76683 Light Sensor Evaluation Software window shown in Figure 4 opens This window is the main interface for demonstrating the capabilities of ISL76683 FIGURE 3 SOFTWARE STARTUP FIGURE 4 ISL76683 LIGHT SENSOR EVALUA...

Страница 4: ... put the ISL76683 into power down mode MEASUREMENT MODE SELECTION The ISL76683 contains two photodiodes Diode 1 is sensitive to both visible and infrared light while Diode 2 is sensitive mostly to infrared light Measurement Mode 1 is Diode 1 only and Measurement Mode 2 is Diode 2 only Measurement Mode 3 is a sequential Mode 1 and Mode 2 with an internal subtract function Diode 1 Diode 2 You can se...

Страница 5: ...n the ISL76683EVAL1Z evaluation board or you can set it to be polled automatically To poll it manually click Sample External Int Pin To poll it automatically click the Poll Interrupt check box to select it With interrupt limits set you can begin collecting data Data is collected within the upper and lower limits you selected In the Interrupt Control dialog in the Interrupt Monitor area black boxes...

Страница 6: ...ding field displays the value of the ADC output coming out of the sensor in accordance with the mode that is engaged Click Stop Data Acquisition to stop data sampling Click Exit to close the entire program In the Scale Max text box enter a maximum value for the scale vertical axis and click Manual Re Scale to re set Click Automatic Re Scale to re scale the vertical axis to an appropriate field of ...

Отзывы: