International Light Technologies ILT1400 Скачать руководство пользователя страница 28

8.8  SPATIAL RESPONSE

8.8.1  Lambertian “Cosine” Response

Lambertian Response is in reference to a particular 

angular response which is proportional to the cosine of the 
incident angle.  In other words, for the angle normal to the 
face of the detector (0 degrees), the magnitude is the cosine 
of 0 degrees, or 1.0 (100%).  As the angle goes off axis and 
becomes parallel to the face of the detector, the reading goes 
to zero, since the cosine of 90 degrees is zero.  At 45 
degrees, the cosine is 0.707, which means that the detector 
should read the rays with 70.7% of the value produce by the 
same rays entering normal to the input device.

The reason this spatial response is necessary for

accurate measurements is that it matches the spatial response 
of a perfect absorbing surface.  Since Irradiance and 
Illuminance are measurements of light falling on a surface, 
the cosine is compatible with these measurements.  An 
analogy of the perfect absorber might be considered as being 
a small hole in a piece of sheet metal placed over a well.  All 
the light that goes in that hole will be absorbed by the deep 
well hole underneath.  None will get reflected back up out of 
the same hole.  If we analyze the effect of a change of angle, 
such as the sun moving from high noon to sunset, we will
see that less light can make it into the hole at sunset, 
because the effective area of the hole is smaller as you view 
it from an oblique angle.  This reduction in area is directly 
proportional to the cosine of the angle normal to this 
surface.  On polar plotting paper, the cosine makes a circle,
which is convenient when comparing the ideal response with 
that of an actual plot.

8.8.2  Field Baffling

There are times when you should restrict the field of view to 
delete oblique angles.  In a lab environment, you may be 
working with a light source on an optical bench. The only 
light of interest is from that source, yet light bounces off the 
people in the room and back to the detector, creating errors 
in the readings.  This means that you are better off to 
restrict the field of view if you know there are no sources to 
be measured at the oblique angles.  This can be done with 
external baffles or with our accessory hood
(H).  Baffles can be made from sheet metal cut to form a 
sharp edged hole in the middle.  A square hole is actually 
better than a round hole, since it is less likely to create 
reflections in a multi-baffle array.  Also, black velvet is 
excellent for dividing off test areas from the rest of the room 
lighting.  If it is necessary to have light travel down a tube, 
you can thread the inside of the tube to reduce the wall 
reflections.  When making Luminance or Radiance 
measurements, it is absolutely necessary to restrict the field 
of view to one that ‘sees’ only an intended test area of a 
reflecting or rear-lit surface.  Baffles can be used to 
implement this kind of measurement without resorting to 
expensive optics.

8.8.3  Narrow Angle (Luminance / Radiance)

As just mentioned, there is a requirement for a narrow 

field of view when making Luminance and Radiance 
measurements.  This can be accomplished by lenses as in our 
Radiance barrel “R” accessory.  In some applications it is 
accomplished by using a telescope where the light is picked 
up from a small spot in the image plane.  This is very nice 
for measuring the brightness of an illuminated segment of an 
alphanumeric display, or for measuring the dot brightness on 
the face of a CRT.  Unfortunately, these systems are very 
expensive.  Another alternative is to use simple lenses to 
image a small portion of a test field onto an aperture which 
has a detector behind it.  This is very effective, especially if 
the source is a repetitive configuration in a production 
situation.  A custom setup can be made to specifically 
measure that on a small emitting surface.  Our “R” barrel has 
a 1.5 degree field of view, with the objective lens being 
about one inch in diameter.  This is effective as long as the 
target is larger than one inch.  The target requirement is 
calculated by multiplying the distance
(measured from the target surface to the front lens of the R 
Barrel, in inches) times 2*tan(2.5/2), plus the 1 inch (25.4 
mm) diameter of the input aperture and some margin for 
error.  If you were 12 inches away, the detector ‘sees’ a 
circular target approximately 1.5 inches in diameter.  Since 
the target should generally be much larger than the field of 
view of the detector, the target should be at least 2 inches in 
diameter when you are 12 inches away.

8.8.4  Uniform Linear Translation Response

The best type of detector setup for uniform receiver 

sensitivity is the input port of an integrating sphere, since 
any off axis beam will still be captured in the chamber.  For 
flux measurements, it is necessary to have this uniformity so 
that small errors in centering the beam do not contribute to 
any error in the measured reading.  Our narrow beam
adapter (H/N/K15) attachment is designed to accept a few 
millimeters of axial misalignment without appreciable 
changes in the reading.  This is necessary to allow for 
non-critical positioning in laser beam measurements.
     The SEL100/FQ/K15 allows the greatest margin of 
error in axial misalignments.  The large, 100 square 
millimeter silicon cell is fitted with a 15 millimeter 
diameter conical aiming aperture with an attenuated “flat” 
response filter.  The user is simply required to aim his 
laser beam inside the aperture for guaranteed linear 
continuity.  Of course, if centering errors are compounded 
by an angular offset, the laser beam may not directly 
strike the silicon cell, causing significant errors in the 
readings.  Centering and normalizing the detector to the 
incoming beam is a simple procedure and should not 
cause any problems

29 

Содержание ILT1400

Страница 1: ...R e v i s e d 02 2 9 1 2 C R M...

Страница 2: ...any form or by any means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests shou...

Страница 3: ...11 3 8 EXPANDED DETECTOR LIST 11 4 OUTPUTS 11 4 1 DISPLAY MESSAGES 11 4 2 TTL UART SERIAL OUTPUT 13 5 OPTICAL CALIBRATION 16 5 1 N I S T TRACEABLE FACTORY CALIBRATION 16 6 PROGRAM MODE OPERATION 16 6...

Страница 4: ...OPTICAL ACCURACY 31 9 3 SIZE AND WEIGHT 31 9 4 ENVIRONMENTAL SPECIFICATIONS 31 10 MAINTENANCE AND REPAIR 32 10 1 PREVENTIVE MAINTENANCE 32 10 2 BATTERY REPLACEMENT 32 10 3 DISASSEMBLY AND PC BOARD RE...

Страница 5: ...continuous wave signal measurement mode for continuous monitoring of a light source The detector memory establishes the detector limits and autoranging units to provide an immediate answer in an easil...

Страница 6: ...adings A detector must be plugged in when the ILT1400 is first turned on for the instrument to properly initialize and interrogate the detector for calibration factors range limitations and units info...

Страница 7: ...he energy reaching the detector 2 2 3 Hold Pressing the HOLD button while in SIGNAL mode will freeze the last reading on the display and the word HOLD will blink in the status register of the LCD To r...

Страница 8: ...s shut down Failure to check the zero is the biggest cause for erroneous readings To check the zero cover the detector or source being measured If the reading is other than zero press the zero button...

Страница 9: ...calibration factors being sure to press hold at the end If this does not successfully restore the calibration factor then sensor will need EEPROM replacement Please contact our service dept for furthe...

Страница 10: ...e 2 50e 8 to 8 75e 2 W cm2 Measurement range 315 390 nm SEL033 F W Flat Radiometric Head Dynamic Range 7 69e 9 to 2 69e 2 W cm Measurement Range 400 1064 nm SEL033 F R Narrow Field Radiance Head Dynam...

Страница 11: ...nm SEL623 SCS695 W IR Hazard Head Dynamic Range 8 57e 5 to 4 14e 1 W cm2 Measurement Range 695 2100 nm 3 6 Additional Photodetectors Below is a list of the additional detectors available for use with...

Страница 12: ...the status register lower right corner of the LCD when the 6 VDC battery voltage drops below 4 95 V The instrument will operate for approximately 54 hours before the BAT LO error status appears The IL...

Страница 13: ...am of data at a rate of one line one reading per second The output is expressed in digital exponential notation The ILT1400 transmits data once every 0 5 seconds PROGRAM MODE ON OFF TO EXIT AUTO SHUT...

Страница 14: ...s a carriage return The twelfth character is a line feed ASCII Decimal Hex Binary LF 10 0a 0 0 0 0 1 0 1 0 CR 13 0d 0 0 0 0 1 1 0 1 43 2b 0 0 1 0 1 0 1 1 45 2d 0 0 1 0 1 1 0 1 46 2e 0 0 1 0 1 1 1 0 0...

Страница 15: ...linking cursor underneath the specific digit that the user wishes to change using the INCREMENT button The cursor moves to the right wrapping around to the first digit on the left after passing the la...

Страница 16: ...hat were not entered will be reset when the instrument is turned off ON OFF button To EXIT the programming mode at any time PROGRAM MODE Quick Reference continued WARNING THE N I S T TRACEABLE CALIBRA...

Страница 17: ...ILY STORED IN RAM AFFECTING THE OPERATION OF THE INSTRUMENT ONLY UNTIL THE POWER IS TURNED OFF 6 2 Disabling the Automatic Shut Off 6 1 7 MENU 7 6 1 8 MENU 8 6 1 9 MENU 9 6 1 10 MENU 10 18 The first m...

Страница 18: ...ity conversions such as lm m2 to lux etc The mathematical conversion can however be accomplished by changing the sensitivity factor 6 4 Changing the Sensitivity Factor WARNING THE N I S T TRACEABLE CA...

Страница 19: ...ange the direction the factor will be incremented either UP to increase or dn to decrease Press the INCREMENT INTEGRATE button to increment the reading You will notice that the reading decreases when...

Страница 20: ...Geist Applied Optics volume 18 number 23 1 December 1979 and also Silicon Photodiode Absolute Spectral Response Self Calibration E F Zalewski and J Geist Applied Optics volume 19 number 8 15 April 19...

Страница 21: ...o modify the circuitry in your ILT1400 Tampering with the electronics will void the warranty If modifications are necessary we can provide schematics of the area to be modified and technical assistanc...

Страница 22: ...ptance angle of the detector Luminance and radiance detectors typically have very narrow acceptance angles 2 5 degrees for measuring the brightness of a uniformly radiating area source When you back a...

Страница 23: ...m units are expressed in Lumens per square foot or foot candles fc One foot candle is equal to 10 76 lux We recommend using lux since they are recognized worldwide When making an illuminance measureme...

Страница 24: ...bilirubin to provide optimal measurement of the therapeutic radiation As with most phototherapy applications it is important to place the detector in the same reference plane as the patient s skin The...

Страница 25: ...pattern for use in photolithography industries such as Printed Circuit Board manufacturing Printing Plate production and XRL140B provide low profile and durability for repeated use in these intense UV...

Страница 26: ...g sphere and the Mean Spherical Candela MSC measurement is a better indication of performance On the other hand if the output from the source is used in one direction then beam intensity measurements...

Страница 27: ...ercent 8 7 REFLECTANCE MEASUREMENTS Reflectance is similar to transmission with a few more complications The ultimate use determines if it is important to measure specular reflectance diffuse reflecta...

Страница 28: ...ded test area of a reflecting or rear lit surface Baffles can be used to implement this kind of measurement without resorting to expensive optics 8 8 3 Narrow Angle Luminance Radiance As just mentione...

Страница 29: ...levels The ILT1400 has the ability to extend measurement time when measuring low light levels so longer exposure times may be required For this reason ILT recommend using the ILT1700 for pulsed light...

Страница 30: ...io and television reception It has been type tested and found to comply with the limits for a class B computing device in accordance with the specifications in subpart J of part 15 of FCC rules which...

Страница 31: ...ws which fasten the back panel of the instrument We recommend using a standard Phillips type screwdriver to avoid damaging the screw heads B Carefully pull the back cover off making sure not to pull o...

Отзывы: