International Light Technologies ILT1400 Скачать руководство пользователя страница 24

For a more empirical calculation of the Effective Dose 

of UVA, simply place the detector in the booth at the level 
of the patient’s skin and put the ILT1400 in ‘INTEGRATE’ 
mode.  The instrument will automatically calculate the 
ongoing accumulated dose in Joules/cm

2

.  Remember to 

press ‘HOLD’ if you wish to terminate or pause the 
integration.  If you hit ‘HOLD’ a second time, the 
instrument will unfreeze the display and return to the 
uninterrupted integration, allowing intermediate readings to 
be taken.  Constant monitoring of fluorescent exposure 
booths is necessary, both to ensure a safe, effective 
ultraviolet dosage and to evaluate when the fluorescent 
lamps need replacement.  As mentioned above, after only
6000 hours of use, a patient will need to be exposed for up 
to three times as long to accumulate the same treatment 
dose.

8.2.4  Phototherapy - UVB

Ultraviolet-B radiation consists of any light in the 

spectral band between 280 and 315 nm.  For many 
phototherapy applications, this “Actinic” radiation can be 
replaced by UVA, since UVB often irritates or damages 
the skin, leading to skin cancer.  For certain skin disorders, 
however, moderate doses of UVB light combined with 
certain drugs can be remarkably beneficial.  Psoriasis, a 
skin irritation condition, can be controlled in this manner. 
Close monitoring of UVB lamp output and doses is 
ESSENTIAL, considering the dangers of overexposure as 
well as the lamp degradation problem mentioned above. 
Also, remember to always wear protective eye wear when 
working with short wavelength sources.

The technique for measuring UVB wavelength light is 

similar to the UVA radiometric irradiance dose measurement 
described above.  The UVB detector should be placed at the 
same reference plane as the patient’s skin.  Be sure to use 
your UVB detector (SEL240/UVB1/TD, for example), not 
your UVA detector, for measuring UVB light.  UVB light is 
much more dangerous than UVA, and your UVA detector 
will not sense any UVB light.  Typical lamps may have a 
large UVA component, leading to misleading results if the 
wrong detector is used.  All International Light 
Technologies detectors are clearly marked with detector type 
and serial number.  If ever in doubt, simply check the 
engraved labels on the individual filter rings to be sure you 
are using the proper detector.

8.2.5

Hyperbilirubinemia Phototherapy - Blue 

Visible

Jaundice is a very common condition requiring medical 
attention in newborns. The yellow coloration of the skin and 
sclera in newborns with jaundice is the result of accumulation 
of unconjugated bilirubin, often referred to as neonatal 
hyperbilirubinemia.

Phototherapy is the primary treatment in neonates with 
unconjugated hyperbilirubinemia. Phototherapy employs blue 
wavelengths of light to alter unconjugated bilirubin in the skin. 
Our Hyperbilirubinemia probe is designed to match the 
wavelength specific photo-oxidation response of bilirubin to 
provide optimal measurement of the therapeutic radiation.

As   with   most   phototherapy   applications,  it   is   important  to 
place   the   detector   in   the   same   reference   plane   as   the 
patient’s skin.    The fact  that  blue light  is visible mislead s 
most   people   into   believing   that   it   is   much   safer   than 
ultraviolet  light.   Excessive  doses  of  blue  light  constitute  a 
significant  health  hazard.    The  U.S.  National  Institute  of 
Occupational   Safety   and   Health   (NIOSH)   has   accurately 
defined  the  “Blue  Hazard”  wavelength  band,  and  the 
maximum allowable daily dosage.  Refer to NIOSH research 
publications for more information on Blue Hazards.

Another important factor to consider when exposing patients 
to  intense  blue  light  is  the  fact  that  many  fluorescent 
hyperbilirubinemia   lamps   emit   a   significant   amount   of 
radiation  in  the  UVA  band  (315  -  400  nm),  which  is  not 
sensed by a bilirubin probe.  We strongly recommend the use 
of  protective  eye  wear,  especially  by  medical  technicians, 
who   may   not   realize   the   ambient   dose   radiation  they  are 
exposed to over time.

8.2.6  ACGIH and NIOSH Actinic UV Hazard

ACGIH has extensively researched the effect of light at 

specific wavelengths on human skin.  They accurately define 
the hazardous effects of ultraviolet light as a function of 
wavelength.  This weighted spectral effect plot is referred to 
as the Actinic hazard function, with the relative hazard 
proportionally weighted from 200 nm to 315 nm (maximum 
hazard at 270 nm).  NIOSH also recommends the use of the 
ACGIH standards to define hazardous working conditions. 
ACGIH updated the Actinic function by logarithmically 
extending this hazard data through the UVA to 400 nm. 
This revision is an important consideration. Although the 
effect of the UVA band is proportionally much less than that 
of the UVB and UVC bands, an intense UVA source poses a 
significant Actinic threat.  We have custom designed our 
Actinic filter using thin film techniques to precisely match 
the new Actinic hazard function curve released by ACGIH, 
out to 400 nm.  This is essential for accurate Effective 
Irradiance measurements.  The ILT1400 combined with 
SEL240/T2ACT displays Effective Watts/cm

2

, providing 

immediate, direct hazard readings when compared to 
Threshold Limit Values.

8.2.7  IES Luckeish and DIN Germicidal / UVC

The germicidal band is precisely defined by IES 

Luckeish and DIN standard curves.  Light in this wavelength 
band acts as a germicide, killing germs and microorganisms. 
Many hospital operating rooms employ mercury arc lamps, 
shining germicidal light inside an open wound as the 
operation takes place to kill environmental germs. Almost 
98% of a Low Pressure Mercury lamp output occurs at 
precisely 254 nm.  The peak of the effective irradiance 
curve, however, is at 265 nm.  Light from a Mercury lamp 
has only 85% of the effect that an equal amount of 265 nm 
light would have.  Our SEL240/T2ACT5 germicidal detector 
weights the effect of all light in this band to its actual 
germicidal effect, regardless of the type of source used.  A 
narrow band detector measuring only light at 254 nm can 
also be used (XRL140T254)

25 

Содержание ILT1400

Страница 1: ...R e v i s e d 02 2 9 1 2 C R M...

Страница 2: ...any form or by any means electronic or mechanical including photocopying recording or any information storage and retrieval system without permission in writing from the copyright owner Requests shou...

Страница 3: ...11 3 8 EXPANDED DETECTOR LIST 11 4 OUTPUTS 11 4 1 DISPLAY MESSAGES 11 4 2 TTL UART SERIAL OUTPUT 13 5 OPTICAL CALIBRATION 16 5 1 N I S T TRACEABLE FACTORY CALIBRATION 16 6 PROGRAM MODE OPERATION 16 6...

Страница 4: ...OPTICAL ACCURACY 31 9 3 SIZE AND WEIGHT 31 9 4 ENVIRONMENTAL SPECIFICATIONS 31 10 MAINTENANCE AND REPAIR 32 10 1 PREVENTIVE MAINTENANCE 32 10 2 BATTERY REPLACEMENT 32 10 3 DISASSEMBLY AND PC BOARD RE...

Страница 5: ...continuous wave signal measurement mode for continuous monitoring of a light source The detector memory establishes the detector limits and autoranging units to provide an immediate answer in an easil...

Страница 6: ...adings A detector must be plugged in when the ILT1400 is first turned on for the instrument to properly initialize and interrogate the detector for calibration factors range limitations and units info...

Страница 7: ...he energy reaching the detector 2 2 3 Hold Pressing the HOLD button while in SIGNAL mode will freeze the last reading on the display and the word HOLD will blink in the status register of the LCD To r...

Страница 8: ...s shut down Failure to check the zero is the biggest cause for erroneous readings To check the zero cover the detector or source being measured If the reading is other than zero press the zero button...

Страница 9: ...calibration factors being sure to press hold at the end If this does not successfully restore the calibration factor then sensor will need EEPROM replacement Please contact our service dept for furthe...

Страница 10: ...e 2 50e 8 to 8 75e 2 W cm2 Measurement range 315 390 nm SEL033 F W Flat Radiometric Head Dynamic Range 7 69e 9 to 2 69e 2 W cm Measurement Range 400 1064 nm SEL033 F R Narrow Field Radiance Head Dynam...

Страница 11: ...nm SEL623 SCS695 W IR Hazard Head Dynamic Range 8 57e 5 to 4 14e 1 W cm2 Measurement Range 695 2100 nm 3 6 Additional Photodetectors Below is a list of the additional detectors available for use with...

Страница 12: ...the status register lower right corner of the LCD when the 6 VDC battery voltage drops below 4 95 V The instrument will operate for approximately 54 hours before the BAT LO error status appears The IL...

Страница 13: ...am of data at a rate of one line one reading per second The output is expressed in digital exponential notation The ILT1400 transmits data once every 0 5 seconds PROGRAM MODE ON OFF TO EXIT AUTO SHUT...

Страница 14: ...s a carriage return The twelfth character is a line feed ASCII Decimal Hex Binary LF 10 0a 0 0 0 0 1 0 1 0 CR 13 0d 0 0 0 0 1 1 0 1 43 2b 0 0 1 0 1 0 1 1 45 2d 0 0 1 0 1 1 0 1 46 2e 0 0 1 0 1 1 1 0 0...

Страница 15: ...linking cursor underneath the specific digit that the user wishes to change using the INCREMENT button The cursor moves to the right wrapping around to the first digit on the left after passing the la...

Страница 16: ...hat were not entered will be reset when the instrument is turned off ON OFF button To EXIT the programming mode at any time PROGRAM MODE Quick Reference continued WARNING THE N I S T TRACEABLE CALIBRA...

Страница 17: ...ILY STORED IN RAM AFFECTING THE OPERATION OF THE INSTRUMENT ONLY UNTIL THE POWER IS TURNED OFF 6 2 Disabling the Automatic Shut Off 6 1 7 MENU 7 6 1 8 MENU 8 6 1 9 MENU 9 6 1 10 MENU 10 18 The first m...

Страница 18: ...ity conversions such as lm m2 to lux etc The mathematical conversion can however be accomplished by changing the sensitivity factor 6 4 Changing the Sensitivity Factor WARNING THE N I S T TRACEABLE CA...

Страница 19: ...ange the direction the factor will be incremented either UP to increase or dn to decrease Press the INCREMENT INTEGRATE button to increment the reading You will notice that the reading decreases when...

Страница 20: ...Geist Applied Optics volume 18 number 23 1 December 1979 and also Silicon Photodiode Absolute Spectral Response Self Calibration E F Zalewski and J Geist Applied Optics volume 19 number 8 15 April 19...

Страница 21: ...o modify the circuitry in your ILT1400 Tampering with the electronics will void the warranty If modifications are necessary we can provide schematics of the area to be modified and technical assistanc...

Страница 22: ...ptance angle of the detector Luminance and radiance detectors typically have very narrow acceptance angles 2 5 degrees for measuring the brightness of a uniformly radiating area source When you back a...

Страница 23: ...m units are expressed in Lumens per square foot or foot candles fc One foot candle is equal to 10 76 lux We recommend using lux since they are recognized worldwide When making an illuminance measureme...

Страница 24: ...bilirubin to provide optimal measurement of the therapeutic radiation As with most phototherapy applications it is important to place the detector in the same reference plane as the patient s skin The...

Страница 25: ...pattern for use in photolithography industries such as Printed Circuit Board manufacturing Printing Plate production and XRL140B provide low profile and durability for repeated use in these intense UV...

Страница 26: ...g sphere and the Mean Spherical Candela MSC measurement is a better indication of performance On the other hand if the output from the source is used in one direction then beam intensity measurements...

Страница 27: ...ercent 8 7 REFLECTANCE MEASUREMENTS Reflectance is similar to transmission with a few more complications The ultimate use determines if it is important to measure specular reflectance diffuse reflecta...

Страница 28: ...ded test area of a reflecting or rear lit surface Baffles can be used to implement this kind of measurement without resorting to expensive optics 8 8 3 Narrow Angle Luminance Radiance As just mentione...

Страница 29: ...levels The ILT1400 has the ability to extend measurement time when measuring low light levels so longer exposure times may be required For this reason ILT recommend using the ILT1700 for pulsed light...

Страница 30: ...io and television reception It has been type tested and found to comply with the limits for a class B computing device in accordance with the specifications in subpart J of part 15 of FCC rules which...

Страница 31: ...ws which fasten the back panel of the instrument We recommend using a standard Phillips type screwdriver to avoid damaging the screw heads B Carefully pull the back cover off making sure not to pull o...

Отзывы: