DATA: Shows the data analysis from when Graph was first pressed, including Min, Max, and 
Average.

EXIT: Returns the display to numerical /non graphing mode and clears the previous graph data 
from the display. 

3. TAKING A MEASUREMENT

The ILT750 sensor is calibrated with the entire surface of the domed diffuser over-filled with 
light. The calibration reference plane is the front surface of the housing. When taking an 
irradiance or band irradiance measurement, the same conditions should be applied. The light 
source to be measured must fully cover the sensor. The sensor should be placed so that the 
front surface is at the treatment location and perpendicular to the source.

The ILT750 offers simple operation. Simply turn on the lamp, press and hold button 1 to turn on 
the meter, place the sensor in the measurement location and view the results on the display.

 3A Measurement Overview:

Long Press button 1 until the meter turns on.

Change the units as needed by pressing button 3.

Turn on the lamp and let it warm up 2 minutes (or longer).

Place the sensor in the measurement location with the front surface of the sensor 
perpendicular to the lamp, at the same distance as the patient will be placed.

The meter will begin providing a readout on the display, updating every ½ second.

Press button 4 to graph and analyze the results as needed.

Long press button 1 to shut off your device, or wait 5 minutes and the meter will shut off 
automatically to conserve battery life.

(Note: the ILT750 meter is a lamp/light source diagnostic tool.  The meter and sensor 
head are not intended to be used while a patient is undergoing treatment or to come in 
contact with the patient.)

4. GRAPHING

The graph mode is designed to test the uniformity or stability of a light’s output for up to 30 
seconds.

The bar graph updates every 1/2 second offering an easy to evaluate, visual response to 
changes in intensity over time. The graph can display up to 60 readings (30 seconds) and will 
begin to scroll forward showing only the most current 60 readings when used for longer 
durations.

If the sensor is moved slowly across a bili-pad/blanket or under a lamp, the ILT750 will map the 

uniformity

 of the output over the target area. If the sensor is held stationary, graph mode can 

verify the 

stability

 of the lamps output.

ILT750-BILI454 Manual Rev A,  Dated  4/22/2021

Page number 5 or 7 

Содержание BiliBlanket ILT750-BILI454

Страница 1: ...ILT750 BILI454 Blue Light Meter Operation Manual ILT750 BILI454 Manual Rev A Dated 4 22 2021 Page number 1 or 7...

Страница 2: ...1 Introduction 3 2 Button Features 3 3 Taking a Measurement 5 3A Measurement Overview 5 4 Graphing 5 5 Battery 6 6 Sensor 6 7 Calibration 7 8 Warranty 7 ILT750 BILI454 Manual Rev A Dated 4 22 2021 Pa...

Страница 3: ...ntimeter per nanometer W cm nm It can be used to measure all types of GE neonatal phototherapy blue light sources including LED light sources illuminated pads and blankets and traditional lamp based l...

Страница 4: ...r with an opaque object wait 2 3 seconds and long press the zero button for 4 sec 3 UNITS Short press to toggle units and calibration factors Though W cm2 nm is becoming a more predominantly used set...

Страница 5: ...tion with the front surface of the sensor perpendicular to the lamp at the same distance as the patient will be placed The meter will begin providing a readout on the display updating every second Pre...

Страница 6: ...The meter is programmed with a 5 minute auto off timer This will help prevent accidental battery drain during use storage or shipping The meter will automatically detect a low battery condition displa...

Страница 7: ...ommends an annual calibration for all equipment however the end user may through their own QC process create their own calibration cycle Service Please be sure to obtain an RMA number on the ILT websi...

Отзывы: